Series S36 Instruction Manual
Chapter 7 Troubleshooting and Repair
800120
7-10
Determine the Fault
Symptom: Output at No Flow
1.
The programming values defined in the
Calibration
menu may not match the
actual hardware being used at the time of measurement, leading to erroneous flow
rate readings. Verify that the programmed values match the hardware that is being
used with the handheld unit.
Symptom: Erratic Output
1.
Mechanical installation may be incorrect. Verify the straight run is adequate as
described in chapter 2 of the manual.
2.
The meter may be reacting to actual changes in the flow stream. The output can
be smoothed using a time constant. The displayed values can be smoothed using
the time constant in the
Display
menu. The analog outputs can be smoothed using
the time constant in the
Output
menu. A time constant of 1 will result in the
change in value reaching 63% of its final value in one second. A time constant of
4 is 22%, 10 is 9.5%, and 50 is 1.9% of the final value in one second. The time
constant equation is shown below (TC= Time Constant).
% 𝑐ℎ𝑎𝑛𝑔𝑒 𝑡𝑜 𝑓𝑖𝑛𝑎𝑙 𝑣𝑎𝑙𝑢𝑒 𝑖𝑛 𝑜𝑛𝑒 𝑠𝑒𝑐𝑜𝑛𝑑 = 100 (1 − 𝑒
(−
1
𝑇𝐶
)
)
Symptom: No Output or Empty Pipe Detected
1.
Verify that there is fluid in the pipe and that there is a full pipe condition. The
flow meter will not read depending on the mounting position of the transducer in
a partially full pipe condition.
2.
Are there air bubbles in the pipe? If so, where are the transducers mounted? If the
transducers are not generating a measurement when mounted on the top/bottom of
the pipe, try moving them so that they are in the 3 O’clock or 9 O’clock position
on the outside of the pipe. This will eliminate interference from entrained air
bubbles traveling in the upper portion of the pipe or solids traveling in the lower
portion of the pipe.
3.
Carefully check all the wiring connections between the fixed mount unit and
transducers. There are four connections that must be correct. Each female
connector has a keyway, and each male connector has a key. These must be
aligned properly when connecting the two for a valid connection to result. If there
are additional wires connected to the terminal blocks on the inside of the
enclosure, ensure all wires are secured by lightly tugging on the insulated part of
the wire to verify it will not pull out. Next, verify that the metal conductors of the
wire are contacting the conductive body of the terminal block.
4.
The velocity cutoff (
Cutoff V)
is set too high. Go to the
First Level of the Hidden
Diagnostics
menu and record the
V
Value. The
Cutoff V
must be set below this
value.
Summary of Contents for SonoPro Professional S36 Series
Page 44: ...Series S36 Instruction Manual Chapter 3 Operation 3 5 800120 Output Menu...
Page 92: ...Series S36 Instruction Manual Appendix A Specifications A 2 800120...
Page 96: ...Series S36 Instruction Manual Appendix C Sound Speed and Pipe Data C 2 800120 Pipe Chart...
Page 97: ...Series S36 Instruction Manual Appendix C Sound Speed and Pipe Data C 3 800120...
Page 98: ...Series S36 Instruction Manual Appendix C Sound Speed and Pipe Data C 4 800120...
Page 99: ...Series S36 Instruction Manual Appendix C Sound Speed and Pipe Data C 5 800120...
Page 100: ...Series S36 Instruction Manual Appendix C Sound Speed and Pipe Data C 6 800120...
Page 101: ...Series S36 Instruction Manual Appendix C Sound Speed and Pipe Data C 7 800120...
Page 102: ...Series S36 Instruction Manual Appendix C Sound Speed and Pipe Data C 8 800120...