S
PECIFICATION
Page 76
© Voltech 2010
Power Analyzer User Manual
45-65Hz
Accuracy
[[(Verr / Vreading) + (Aerr/Areading)] * VA
Reading]
Note1,3
DC
Accuracy
0.2% of R 0.4% of range +4mVA
VAr
RANGES
1 VAr to 90kVAr
Frequency range
10Hz to 1MHz
RMS
Accuracy
0.2% of R 0.1% of range +4mVAr
+((0.05 / (1- PF)) * F)% of reading
45-65Hz
Accuracy
[[(Verr / Vreading) + (Aerr/Areading)] * VAr
Reading] + (0.1 / (1
– PF))% of VAr
reading
Note1,3
DC
Accuracy
0.2% of R 0.4% of range +4mVAr
POWER FACTOR
Range
+- 0.000 to 1.000
Accuracy
+-(0.002+-(0.001 / PF) * F)
FREQUENCY
Range
DC and 10Hz to 1MHz
(Maximum 22kHz when frequency source
is set to current)
Accuracy
0.1%
VOLTAGE CREST
FACTOR
RANGE
1.00 to 20.0
Accuracy
%Vpk error + %Vrms error
CURRENT CREST
FACTOR
RANGE
1.00 to 20.0
Accuracy
%Apk error + %Arms error
PEAK INRUSH
CURRENT
RANGE
100Apk
Accuracy
2% of range +/- 20mA