Select
20M
in the right menu. The bandwidth is limited to 20 MHz. The frequencies
above 20MHz will be rejected.
6.
To set the input impedance
To reduce the circuit load caused by the interaction of the oscilloscope and the circuit to
be tested, you can set the input impedance.
Taking the Channel 1 for example, the operation steps are shown as below:
(1)
Push the
CH1
button to show CH1 SETUP menu.
(2)
Select
Input IMP
in the bottom menu, push to select
1MΩ
or
50Ω
.
1MΩ
: The input impedance of the oscilloscope is rather high and the current flowing
into the oscilloscope from the circuit under test can be ignored.
50Ω
:
Match the oscilloscope with a device with 50 Ω output impedance. The max
input voltage cannot exceed 5 Vrms.
Use Mathematical Manipulation Function
The
Mathematical Manipulation
function is
used to show the results of the addition,
multiplication, division and subtraction operations between two channels, the FFT
operation for a channel, advanced math feature including Intg, Diff, Sqrt, user defined
function, and digital filter. Press the
Math
button to display the menu on the bottom.
The
Waveform Calculation menu:
Function Menu
Setting
Description
Dual Wfm
Math
Factor1
CH1
CH2
Select the signal source of the factor1
Sign
+ - * /
Select the sign of
mathematical manipulation
Factor2
CH1
CH2
Select the signal source of the factor2
Vertical
(div)
Turn the
M
knob to adjust the vertical position of the Math
waveform
Vertical
(V/div)
Turn the
M
knob to adjust the vertical division of the Math
waveform
FFT
Source
CH1
CH2
Select CH1 as FFT source.
Select CH2 as FFT source.