4.Advanced User Guidebook
83
Measure
Only if the waveform channel is in the ON state, the measurement can be performed.
The automatic measurement can not be performed in the following situation: 1) On
the saved waveform. 2) On the Dual Wfm Math waveform. 3) On the Video trigger
mode.
On the Scan format, period and frequency can not be measured.
Measure the period, the frequency of the CH1, following the steps below:
1.
Push the
Measure
button to show the automatic measurement function menu.
2.
Select
Add
in the bottom menu.
3.
In the right menu, select
CH1
in the
Source
menu item.
4.
In the left Type menu, turn the
M
knob to select
Period
.
5.
In the right menu, select
Add
. The period type is added.
6.
In the left Type menu, turn the
M
knob to select
Frequency
.
7.
In the right menu, select
Add
. The frequency type is added.
The measured value will be displayed at the bottom left of the screen automatically
(see
Figure 4-14 Automatic measurement
The automatic measurement of voltage parameters
The oscilloscopes provide automatic voltage measurements including Mean, PK-PK,
RMS, Max, Min, Vtop, Vbase, Vamp, OverShoot, PreShoot, Cycle RMS, and
Cursor RMS.
below shows a pulse with some of the voltage
measurement points.