c) Measurement tolerances
Accuracy: ± (% of r count)
The accuracy is valid for one year at:
• Ambient temperature: +3 °C (± 5 °C), ≤75 %, RH non-condensing
• Operating temperature*: 18 to 28 °C (±1 °C)
*A temperature coefficient applies outside of this temperature range: +0.1 x (specified accuracy)
/ °C.
The measurement can be impaired when the device is operated within a high-frequency
electromagnetic field.
d) Continuity ( ) and diode ( ) test
Range
Resolution
Remark
0.1 Ω
• Open circuit: Resistance >100 Ω, no beep.
• Circuit with a good connection: Resistance ≤10 Ω, con-
secutive beeps.
1 mV
• Open circuit voltage: Approx. 2.2 V
• Silicon PN junction voltage: Appox. 0.5 to 0.8 V
Overload protection: 600 V
e) DC voltage measurement
Range
Resolution
Accuracy
200.0 mV
0.1 mV
± (0.5 % + 8)
2.000 V
0.001 V
20.00 V
0.01 V
200.0 V
0.1 V
600 V
1 V
± (0.8 % + 7)
• Input impedance: ≥100 MΩ for mV range (short circuit allows ≤5 digits), approx 10 MΩ
for other ranges.
• Input voltage: max. 600 V
54