d) Continuity test ( )
Make sure all circuit parts, circuits, components, and other objects of measurement
are disconnected from the voltage supply and fully discharged.
1. Set the function dial to .
2. “ ”will show on the display.
3. Connect the test leads to the terminals:
- Red test lead to the
terminal.
- Black test lead to the COM terminal.
4. Connect the test lead probes to the correct
test points in the circuit.
5. A resistance of ≤ 10 Ω is considered conti-
nuity, and a beep will sound.
- The measuring range is ≤ 200 Ω.
- “OL” will appear on the display if the
maximum range is exceeded or if the
circuit is open.
6. When finished measuring, disconnect the
test leads and switch the power OFF.
45