Use the “Hz%” button (11) to switch the display to frequency (Hz) and pulse length
(%). Each press switches the function. Autorange is deactivated. This enables monito-
ring of possible ripple signal (superimposed frequency). In case of pure direct current,
the frequency and % displays are “zero”. Without a measured signal, phantom values
may be displayed. They disappear when a direct current is measured.
Proceed as follows to measure AC voltages “A-AC”:
• Turn the DMM on and select measuring function “A”.
• The table shows the different measuring functions and possible measuring ranges. Select
your measuring range and the respective measuring jacks. The display indicates “DC”.
Measuring function
Measuring range
Measuring jacks
µA
<4000 µA
COM + mAµA
mA
4mA – 399 mA
COM + mAµA
A
400 mA – 10 A
COM + 10A
• Insert the red measuring line into the µA/mA or 10A measuring jack. Plug the black measuring
line into the COM socket.
• Press “MODE” to switch to the AC measuring range. “AC” appears in the display. Pressing
this button again, takes you back etc.
• Now connect the two measuring prods in series to the object to be measured (power circuit,
switch etc.). The measured value is indicated on the display The connection principle cor-
responds to DC measurement.
• After measuring, remove the measuring lines from the measured object and turn the DMM off.
Use the “Hz%” button (11) to switch the display to frequency (Hz) and pulse length
(%). Each press switches the function. Autorange is deactivated.
60