40
11. Remote mode
Based on the instruction, turn on the PC control functionality, set the parameter of serial interface on PC and send the
protocol command to control the calibrator.
Download the software from www.conrad.com/downloads.
12. Advanced application
a) Percentage
When the calibrator is in output mode, short press
,
,
,
to rapidly output percentage value
accordingly, the 0% or 100% value of each output functionality is as follows:
Output Functionality
0% value
100% value
Voltage
0 V
10 V
Millivolt
0 mV
100 mV
Resistance 400 Ω
0 Ω
400 Ω
Resistance 4000 Ω
0 Ω
4000 Ω
Thermocouple J
-200 °C
1200 °C
Thermocouple K
-200 °C
1372 °C
Thermocouple T
-250 °C
400 °C
Thermocouple E
-200 °C
1000 °C
Thermocouple R
0 °C
1767 °C
Thermocouple S
0 °C
1767 °C
Thermocouple B
600 °C
1820 °C
Thermocouple L
-200 °C
900 °C
Thermocouple U
-200 °C
400 °C
Pt100-385
-200 °C
850 °C
Pt100-392
-200 °C
630 °C
Pt100-JIS
-200 °C
630 °C
Pt200-385
-200 °C
630 °C
Pt500-385
-200 °C
500 °C
Pt1000-385
-200 °C
650 °C
Cu10
-100 °C
260 °C
Cu50
-50 °C
150 °C
Cu100
-50 °C
150 °C
Ni120
-80 °C
260 °C
Summary of Contents for 2389432