d) Fuse Replacement / Product Disposal

•  Prior to fuse replacement, ensure that device is disconnected from external voltage supply and the other connected 

instruments (such as UUT, control instruments, etc.)

•  Only use fuses as described in the technical data section!
•  Using auxiliary fuses, in particular short-circuiting fuse holders is prohibited and can cause instrument destruction 

or serious bodily injury of operator.

•  Switch off the instrument. Disconnect test leads.

 - Loosen the screws on the instrument rear.
 - Lift the case cover.
 - Remove the defect fuse.
 - Insert new fuse.
 - Replace the case cover and retighten the screws.
 -

Ceramic fuse F1/F2 : 16A, 250V, fast-blow (5x 20 mm), breaking capacity ≥500A

•  General warning: if the battery/fuse compartment is open, device must not be used! 

Product Disposal

•  This product must not be disposed of in household waste.
•  Batteries must be removed and disposed separately from the product.

Summary of Contents for 2362387

Page 1: ...Operating Instructions Device tester GT 6000 v2 Item No 2362387 Page 2 34 ...

Page 2: ... devices of protection class 3 safety low voltage 17 d Lead test testing of power extensions cables cable reels and multi plugs 19 e Testing passive and active of three phase devices 22 f Testing of RCDs 23 10 Memory function 26 a Saving measurement data 26 b Retrieving measurements 26 c Measurement data deleting 26 d Transferring measurements data 26 e Software and driver installation 27 f Settin...

Page 3: ...n mark in a triangle indicates important notes in these operating instructions that must be observed strictly The arrow symbol indicates that special advice and notes on operation are provided Conformity symbol the instrument complies with the valid directives It complies with the EMC Directive 2014 30 EU Standards EN 61326 1 is fulfilled It also complies with the Low Voltage Directive 2014 35 EU ...

Page 4: ...e electrical devices testing with the help of optional measurement adapters Voltage test on grounded sockets Cables extension cables cable distributions and IEC power cables testing Tester GT 6000 v2 is characterized by the following features Test results of GT 6000 v2 are compared with preset values of limits and based on that pass fail information is given to user Preset values of limits and mor...

Page 5: ...ow the instructions on the website 5 Optional accessories Passive measurement adapter Single and three phase loads adapter are intended for RPE RISO and IEA measurements without mains voltage dependent switching device Commercially available measurement adapters with the following connections can be used 16 A CEE connector 5 pin L1 L2 L3 are bridged earthing contact connector Active measurement ad...

Page 6: ...for electrical systems and equipment must be strictly always met The respective accident prevention regulations established by the professional associations are to be strictly always enforced regarding body protection in the event of danger of burns In order to avoid electrical shock the valid safety and VDE regulations regarding excessive contact voltages must receive utmost attention when workin...

Page 7: ...rating instrument and measurements battery compartment must be closed Test leads or accessories must be original During measurements touching of measuring points and measuring probes tips outside of the marked handling area is forbidden The instrument may only be used under those conditions and for those purposes for which it was conceived For this reason in particular the safety references the te...

Page 8: ...8 8 Operation elements and connections ...

Page 9: ...on resistance measurement pushbutton for setting the test voltage 250 V DC or 500 V DC 6 Pushbutton for testing RCCB RCD breakers 7 Pushbutton for testing 3 phase devices with optional 3 phase adapter 8 Display 9 4 mm test socket to be used with test lead with alligator clip 10 IEC 60320 C14 connector for connecting the IEC low power cable C13 or for connecting optional measurement 3 phase adapter...

Page 10: ... selected protective conductor resistance differential and touch leakage current measurements Calibration mode only available in service center USB connection STORE saving record in PAT memory OPEN recall saved record from PAT memory Battery is low battery replacement is required Battery is empty PAT will soon Turn Off ...

Page 11: ...t Differential or Touch Leakage Current IDIFF IDIR Test Substitute Leakage Current IALT Test RCD Test Portable RCD Test RCD Test 180 initial polarity RCD Test 0 initial polarity Volt Voltage measurement Ω RPE Test Mega Ω RISO Test RISO Test Voltage 500V RISO Test Voltage 250V ...

Page 12: ...le Test 3 Phase 16A CEE Adapter Note Backlight function white LEDs is activated deactivated by short press to the pushbutton 15 If not deactivated by the user backlight will turn off automatically after approx 1 minute Additionally if the test result is PASS it will be indicated by turning on the green LEDs backlight Otherwise when test result is FAIL the red LEDs backlight is turned on ...

Page 13: ... the device on or off press and hold the button for approx 1 second The device will turn on or off with a beep To avoid electrical shock the valid safety measures and VDE directives strictly must be met concerning excessive contact voltage when working with voltages exceeding 120 V 60 V DC or 50 V 25 V rms AC The values in brackets are valid for limited areas such as e g medicine agriculture Conne...

Page 14: ...tests the device under test during operation connect the mains cable with the built in connector 11 on the GT 6000 v2 and the safety plug to an earthed socket 230 V 50 Hz 16 A Switch on the DUT Device Under Test Start the automatic test by short pressing pushbutton 2 First test is the measurement of the protective conductor resistance RPE Earth continuity Note Extended version of this test is star...

Page 15: ...ment will be performed again with reverse polarity The higher value of both measurements is shown The insulation resistance test is started after RPE test passed successfully If LO LOAD appears on the display should be checked if DUT is switched on If the DUT is on but the load RL N 6 kΩ press the button 2 to proceed with the test sequence Very high load RL N 14 Ω ILOAD 16 A from the DUT is presen...

Page 16: ... DUT should be switched on and active Step 1 of 2 After a measurement time of 5 seconds the polarity mains is reversed after 5 sec and the protective conductor current is measured with the reverse mains voltage L N N L The higher value of both measurements is shown Step 2 of 2 Tick symbol will be displayed beside the ILEAK value if the leakage ILEAK is less than limit PASS will be shown on the dis...

Page 17: ... mains voltage 230 V AC Appliance must be placed on insulated surface If that is not the case possible leakage currents to earth will disrupt the measurement and accuracy c Testing devices of protection class 2 insulated and devices of protection class 3 safety low voltage This function is used to test devices without protective conductor and with touchable conductive parts Plug the DUT Device Und...

Page 18: ...ircuit press button 3 to proceed with the test sequence Beside the RISO value will be shown a tick symbol f the RISO is higher than the permissible minimum value Contact current measuring using the direct measurement method Touch Leakage Current After the RISO measurement GT 6000 v2 stops the test sequence and prompts the user to switch the mains voltage from 230 V to the test socket ILEAK symbol ...

Page 19: ... Option Substitute Leakage Current Battery operation without mains supply Appliance is connected to the GT 6000 v2 like previously described above but without connecting the mains voltage If measured IEA using equivalent leakage current measurement method is lower than the permissible limit value a tick symbol will be displayed beside IEA value PASS will be shown on the display if whole test is su...

Page 20: ...limit need to press the button 2 In that case a tick symbol will be displayed beside the RPE value and the test can continue If main test person decides to RPE value rate like not good above limit need to press the button 3 FAIL will be displayed and test is stopped After RPE test is passed the RISO is measured automatically Tick symbol is displayed next to the RISO symbol if value is above limit ...

Page 21: ...ass 1 will start RPE will be first measured The rest of testing is the same as described above for testing of IEC power cable Testing three phase cable Connect the three phase cable to the test socket 1 using the optional passive measurement adapter Insert the 4 mm safety connector of the test lead with alligator clip into the 4 mm safety socket 9 and connection to the PE on CEE connector of the t...

Page 22: ...e 3 Active testing For active testing adapter 16 A CEE 5 pin option in GT 6000 v2 kit is necessary Adapter is active electronics inside Connect the CEE connector on the device under test to the CEE connector on the measurement adapter and then connect the CEE connector on the measurement adapter to a protected mains supply 3x 400 V N PE 50 Hz 16A The measurement signal cable IEC of the 16 A CEE ad...

Page 23: ...ollowing warning is displayed is warning which can be seen on picture on the right side DUT must be placed on insulated surface If that is not the case possible leakage currents to earth will disrupt the measurement and accuracy f Testing of RCDs 30 mA residual current circuit breakers RCCBs RCDs The GT 6000 v2 supports tripping time measurement of permanently installed RCCBs RCDs and portable RCC...

Page 24: ...ched on The RCCB RCD is tripped by fault current of 30mA generated by GT 6000 v2 with both 0 and 180 polarity The tripping times are measured Tick symbol is displayed next to the tripping time value if the tripping time is less than the limit 200 ms If tripping time caused by the single rated fault current is inside limit GT 6000 v2 will generate 150 mA fault current with 0 or 180 initial polarity...

Page 25: ... the display safety plug in the socket of the RCCB PRCD need to be rotated by 180 By pressing the button 6 the test will be restarted RCCB PRCD should be switched back on each time when rESEt symbol is displayed The portable RCCB PRCD is tripped by fault current of 30 mA with 0 or 180 initial polarity The tripping times caused by such single rated fault current are measured Tick symbol is displaye...

Page 26: ...memory space the save button 13 switches back to the previous memory space c Measurement data deleting Press the folder button 14 to retrieve the saved measured values with the corresponding memory space number The RECALL symbol appears on the display To clear the entire measurement memory press and hold the save button 13 and the folder button 14 at the same time until the counter value is reset ...

Page 27: ...re shortcut will be placed on the desktop Connect the USB C data cable to the USB C connection 12 on the GT 6000 v2 Connect the USB A connector to an unused USB interface on your computer Switch the GT 6000 v2 off To activate the USB interface press and hold the save button 13 and the On Off button 15 at the same time The device switches on and activates the interface The USB plug symbol and USB C...

Page 28: ... 15 at the same time The device switches on and activates the setting The date time format is displayed as follows DD MM Day 1 31 Month 1 12 YYYY Year HH mm Hours 0 23 minutes 0 59 SS Seconds 0 59 not adjustable Press the button 2 to select a date time field Pressing the button moves one field ahead each time When the field flashes the value for this field can be set Press the save button 13 to de...

Page 29: ...nd a calibration interval of one year c Battery Replacement and Disposal Prior to battery replacement disconnect the instrument from any connected test leads Only use batteries as described in the technical data section Switch off instrument Disconnect test leads Loosen the screws on the instrument rear not to be completely removed Lift the case cover Remove discharged batteries Insert six new bat...

Page 30: ...ause instrument destruction or serious bodily injury of operator Switch off the instrument Disconnect test leads Loosen the screws on the instrument rear Lift the case cover Remove the defect fuse Insert new fuse Replace the case cover and retighten the screws Ceramic fuse F1 F2 16A 250V fast blow 5x 20 mm breaking capacity 500A General warning if the battery fuse compartment is open device must n...

Page 31: ...277 x 124 x 68 mm Weight approx 1030 g Ambient conditions Operation Temperature 0 to 30 C 0 80 rel humidity Operation Temperature 31 C to 40 C 0 75 rel humidity Storage Temperature 25 C to 65 C 0 80 rel humidity Overload Protection Fuse F1 F2 F 16 A 250 V Ceramic 5x 20 mm Auto Power Off function 1 minute Battery life approx 2500 tests with full batteries Technical Data refer to 23 C 5 C at 80 rel ...

Page 32: ...9 MΩ 0 1 MΩ 5 2 Test voltages 250 V DC or 500 V DC 20 0 Measuring current 1 mA 2 mA at 2 kΩ Factory Set Pass Fail Preset limits 1 MΩ protection class 1 2 MΩ protection class 2 Substitute Leakage Current Protective conductor and contact current equivalent leakage current measurement method Measurement range Resolution Accuracy 0 25 mA 19 99 mA 0 01 mA 5 2 Test voltage 40 V AC 50 Hz Test current 10 ...

Page 33: ... Current Contact current direct measurement method Measurement range Resolution Accuracy 0 1 mA 1 99 mA 0 01 mA 5 2 Test voltage 230 V 10 Rated current 16 A Max measurement time 30 seconds Factory Set Pass Fail Preset limit 0 5 mA protection class 2 Protected against external voltages Max 276 V For non sinusoidal power supply an additional error must be taken into account Crest factor of 1 4 to 2 ...

Page 34: ... current direct measurement method with optional measurement adapters Measurement range Resolution Accuracy 0 25 mA 9 99 mA 0 01 mA 5 2 Test voltage 3x 400 V 10 Rated current 16 A Factory Set Pass Fail Preset limit 3 5 mA Voltage measurement Voltage measurement at external earthed socket Measurement range Resolution Accuracy 5 V 270 V AC 1 V 5 2 Measured and display Voltage between L and N L and P...

Page 35: ...ved Reproduction by any method e g photocopy microfilming or the capture in electronic data processing systems require the prior written approval by the editor Reprinting also in part is prohibited This publication represent the technical status at the time of printing Copyright 2021 by Conrad Electronic SE 2362387_v1_0521_02_mxs_m_en ...