background image

8. MAINTENANCE

As with any silage inoculant applicator, the tank and tubing should be cleaned at frequent intervals to prevent 
microorganisms (inoculant or contaminants) growing inside the tank or tubing.  Ideally the whole system 
should be washed with clean water between jobs, taking care to clean under the tank rim and flushing the 
tubing well. If microbial growth does occur the tank should be scrubbed with hot, soapy water then the whole 
system flushed thoroughly with clean water so there is no residual soap.  

Draining inoculant from the tank for refrigeration

: Use ‘reverse cleaning mode’ to flush any inoculant in the system back 

into the tank then drain the tank contents into suitable clean containers, eg original bottles. 

Flushing the system:

 

Rinse the inside of the tank thoroughly with clean water and drain off, taking care to clean under 

the rim. Close the drain tap and put some more clean water into tank, turn the 3-way valve to the tubing outlet and run 
the pump in ‘forward cleaning mode’ to flush the system. You may need more than one cycle. Open the tank drain to 
release any water remaining in the tank.

Pump

: Regularly check the pump and inside of the pump box are clean and dry.

Out of use:

 If the pump is not to be used for a prolonged period, remove the rubber tube and holder from inside the pump 

to prevent the tube becoming constricted due to prolonged compression at the contact points (see below). Alternatively, 
once a week remove the clear pump cover and red disc and manually turn the central hub to a new position.

Removing the tube holder from the pump:

 

Remove the clear cover and the red disc then slide out the orange tube holder. 

Turn the centre hub clockwise until it resembles the letter ‘D’. Gently ease the tube out at the left-hand side then rotate 
the central hub clockwise to gradually release the rest. 

Re-inserting the tube holder into the pump: 

Turn the centre hub so it resembles the letter ‘D’ then slide in the orange 

plastic holder. Press the tube  down into place on the left-hand side then gently press the rest of the tube in place while 
rotating the hub clockwise. Replace the red disc, taking note of the alignment underneath, and the plastic cover.

Replacing the pump tube:

 

Make sure the tube is not twisted – the two red marks on the tube should both be facing the 

same way when attached.

End of season:

 

1) Wash the whole system thoroughly with clean water; 2) Leave tank drain open during storage; 

3) Check the pump and inside of the pump box are clean and dry and remove the pump tube; 4) Check the condition of 
the nozzle.

Start of season:

 

1) Check all tubing for damage; 2) We recommend you fit a new pump tube – a spare is supplied with 

the kit; 3) Check the condition of the nozzle and make sure it is correctly positioned in the chute; 4) Run some water 
through the system and test the control box and pump are both functioning correctly.

-7-

Summary of Contents for Ecosyl ECOSYLER 08

Page 1: ...ting mode 4 5 Counter 4 6 Alerts 4 6 1 Low Level 4 6 2 Pump 4 7 Control Box Menu 4 7 1 View the counter reading Counter tonnes or litres 4 7 2 Reset the counter ctr clr CTR 0 6 7 3 Application rate Do...

Page 2: ...et supplied 2 3 Nozzle The nozzle should be placed in the lower side of the chute approximately midway along its length eg fit into the centre of a convenient cleaning flap Drill a 14mm diameter hole...

Page 3: ...ches to Cleaning mode forward or twice 2 rapid clicks to Cleaning mode reverse Pressing the key takes you into the menu where you can reset the counter and change various applicator parameters see sec...

Page 4: ...both of which cause the Alert LED on the control box to flash and an audible warning which can be cancelled by pressing the key The display will return to normal but the red LED continues to flash un...

Page 5: ...NB Before returning to harvest mode you must confirm all 3 digits even if only one changed t h or l h rate of harvest or additive use Counter accesses the counter t or l total tonnes harvested or lit...

Page 6: ...e control box 8 Collect at least 500ml 4 minutes then stop the pump by pressing the key 9 View the counter reading in litres 10 Divide the amount of inoculant collected by the counter reading the coun...

Page 7: ...becoming constricted due to prolonged compression at the contact points see below Alternatively once a week remove the clear pump cover and red disc and manually turn the central hub to a new positio...

Page 8: ...for constrictions or breaks in the tubing and blockages including in the pump Control box power light on but nothing showing on LCD screen Try switching the power source off and on again ie switch ha...

Page 9: ...l Box 7 Tank 8 Pump Housing Cover 9 Pump Housing 10 Pump 11 6mm Hose Tail 12 Quick Connector A 13 Quick Connector B 14 Nozzle Item No Description 15 Quick Connector A 16 Quick Connector B 17 10mm Hose...

Reviews: