Installation Instructions
© BTG 2020
User Manual CPM II2066en Rev. C
21
HCM-8010 Connections
NOTE!
The functions of the connections for each instrument type can be found in the
connection tables for HCM-8010 in the appendix of this manual
P1
P3
P2
P4
P6
P5
P7
P9
P8
Fig 13 HART Communication
Module HCM-8010
P3
HCM-8010
P2
P1 P4
P5
P6
P7
P8
P9
Digital In 2
Digital In 1
Digital In 3
Digital Out
Analog In 4-20 mA
Active Analog Out 4-20 mA
Active Analog Out 2 (4-20 mA)
Active Analog Out 3 (4-20 mA)
Active Analog Out 4 (4-20 mA)
Active Analog Out 5 (4-20 mA)
Digital In 4
Digital Out 2
Solenoid Valve 4
Solenoid Valve 5
Transmitter
cable
Cable shield
SFM-8000
DI
3
P1
DO
DI
1
P2
DI
2
AO
1
P3
AI
1
AO
2
P4
AO
3
AO
4
P5
AO
5
AI
P6
23
4
DI
4
P7
DO
2
SV
4
P8
SV
5
SV
P9
12
3
+
-
LOAD
+24 V
+24 V
+24 V
+24 V
+24 V
Analog In 2 (4-20 mA)
Analog In 3 (4-20 mA)
Analog In 4 (4-20 mA)
24 V
+24 V
Solenoid Valve 1
24 V
Solenoid Valve 2
Solenoid Valve 3
+
-
LOAD
+24 V
Fig 14 Connection of
HCM-8010 cables
CAUTION
Always connect
the cable shield!
CAUTION!
The cable shield must always
be connected to the upper
connection point on the shield
filter module, and kept
separated from the transmitter
cable shield.
NOTE!
For HCM-8010, only active
analog output is available.
Summary of Contents for BTG CPM-1300
Page 1: ...User Manual CPM Communication Platform...
Page 18: ...Product Introduction 14 User Manual CPM PI2066en Rev C BTG 2020...
Page 34: ...Service Instructions 30 User Manual CPM SI2066en Rev A BTG 2020...
Page 50: ...Parts List 46 User Manual CPM PL2066en Rev A BTG 2020...
Page 60: ...Appendix 56 User Manual CPM AX2066en Rev C BTG 2020...
Page 61: ......
Page 62: ...BTG Instruments AB P O Box 602 SE 661 29 S ffle Sweden Phone 46 533 426 00 www btg com...