VocalMist
10
PAGE
|
4. Your nebulizer will power off automatically after 10 minutes. If you need to continue use, press
the Power Button again. Make sure there is enough liquid in the saline cup.
5. After nebulization, press the Power Button to turn off your nebulizer. Remove cap and pour out
any leftover liquid and do not reuse. Leave cap off and allow the saline cup to dry completely
after each use.
Note:
If not held upright, liquid or condensation may become coagulated around the spray connector
and mesh disc and cause the device to stop producing mist. Should nebulization stop, remove
the nozzle and other accessories. Shake the device lightly to dislodge liquid buildup. If the
problem persists, use a clean medical gauze to wipe residue. To prevent damage, do NOT touch
the metal disc.
Note:
Any disinfectant residue should be thoroughly cleaned before resuming use.
Cleaning and Disinfection
1. AFTER EACH USE
♦
Remove the cap and lightly shake any remaining saline solution from the cup. Leave the cap off
and allow the saline cup to dry completely before replacing the cap
♦
The Mask and Mouthpiece may be cleaned with disinfectant or soap and water. They are
dishwasher safe