111
Boot loader
Symptom
Action
Cause
After selecting
(F1)-
Retry to load
application
the
message
Error Bad
program file
appears
Press
(F3)-Load
program from SD card
and follow the steps
described in
Failsafe
Update
Procedure
section in
chapter
Instrument
overview
This occurs when the program file stored in the
internal memory is not valid. If the procedure
fails, there is an internal storage memory
problem.
After selecting
(F3)-
Load from SD Card
the
message
Error Cannot
mount SD card
appears
Put the instrument OFF
and connect it to a PC.
This message is generic for the SD card
controller and appears when either:
- The controller is defective
- The card is defective
- The card is not properly formatted.
The instrument should behave as a standard
mass storage device when connected to the PC.
If the device is not published or appears as an
unknown device, there is an SD card controller
error.Pleace contact a VMI service center.
If the device is published (appears in Control
panel) but cannot be handled as a disk, it is a
problem with the SD card, which is either not
formatted or defective.
Regarding format parameters, read
Format
section in chapter
Instrument overview.
Message:
”Error:
Cannot find SYSTEM
folder”
The problem is due to a
bad format of the card
or a card error.
The problem can also
be due to reason that
the required file or
folder may not be
present on the card.
Put the instrument OFF
and connect it to a PC.
If it is no hardware fault you should use the
operating system utilities to determine and
solve the problem.
The messages are self explaining and will guide
you to solve the problem.
Regarding format, please read
Format
section in
chapter
Instrument overview
.
Regarding files, please read
Files
section in
chapter
Instrument overview
Message:
”Error:
Cannot find any
program file”
Message:
”Error:
Cannot open file”
Message:
”Error: Bad
header CRC”
Message:
”Error:
Incompatible program
file”