- 17 -
5.7 Safety devices
Safety element
responds when
Signal
Assistance
Overheating protection
Heating block temperature
exceeds the maximum ad-
justable set-point of
130/210°C by more than
20°C
see chapter
“Malfunction”
Switch off unit.
Failure corrective action only
through authorised person
or manufacturer
Unit fuse
electric failure occurs
No unit functions
Failure corrective action
through authorised person
or manufacturer, replace
fuse
Controller fuse
electric failure of the control-
ler
No unit functions
Failure corrective action
through authorised person
or manufacturer, replace
fuse
5.8 8 Conformity
Model
Metalblock-Thermostats
HP-Thermostat
Type HP-1V... Type HP-2V... Type HP-1F... Type HP-2F...
EC-Thermostat
Type EC-1V... Type EC-2V...
BIO-Thermostat
Type BIO-1V... Type BIO-2V... Type BIO-1F... Type BIO-2F...
LS-Thermostat
Type LS-1V... Type LS-2V...
QS-Thermostat
Type QS-1V... Type QS-1F...
Guidelines or
designation of standards
Laboratory units to heat materials
Standards
DIN EN 61010-1 (VDE 0411 Part1)
DIN EN 61010-2-010 (VDE
5.9 9 Normal supplies
Metalblock-Thermostat
Mains cable
Spare fuses
interchangeable heating blocks with holders if ordered
operating manual
Special accessories:
Polycarbonate-covers
Safety hoods made of polycarbonate
Summary of Contents for BIO-1F