In order to reassembly the parapodium follow these instructions: 

 

a)

 

Put the corset assembly back (pos.1, pict.4) on the basis assembly (pos.2, pict.4), 

b)

 

Adjust the previous height of the mounting of the corset assembly 

(1), 

 

c)

 

Screw the screws with two pads again (pos.3, pict.4), 

d)

 

Check the condition of equal dimensions X and Y (X=Y). 

 
 
 
 
 
 

Y

 

 
 
 
 
 
 
 
 
 

 

 
 
 

Pict.3 

 
 

2.6

 

Storing, Cleaning And Maintenance Of The Parapodium. 

 

The  Dynamic  Parapodium  is  a  mechanical  device,  which  framework  is  made  of  steel  coated  with  lacquer  and 

chromium according to PN-93/C-81515 and PN-83/H-97006. The soft elements are made of foam-filled leather (ecological 
leather) or velour upholstery. 

 

Parapodium,  as  any  medical  device,  should  be  cleaned  regularly  and  used  according  to  the  manufacturer’s 

instructions. 

 

Recommendations concerning storing 

The device should be stored in a dry, air-conditioned room, in which the relative humidity is not higher than 80%. 

 

Recommendations concerning cleaning and maintenance: 

 

 

Lacquer  coated  surfaces  should  be  cleaned  with  a  damp  piece  of  cloth.  Adding  mild  detergents  used  to  clean 
household appliances is acceptable.

 

 

Leather upholstery should be cleaned and conserved with special agents used to clean leather clothing

 

 

 

Velour upholstery should be washed at 40°C with water containing mild detergents used to wash underwear and 
coloured clothing.

 

 

 

Metal  surfaces  should  be  cleaned  with  alcohol.  The  case  of  cutting  the  metal  layer  should  be  reported  to  the 
service point and a safety layer should be put on. Moreover, any trace of corrosion should also be reported.

 

 

In case of using the parapodium In hospitals and clinics by many patients, and their regulations do not contradict it 
is recommended to disinfect the handles (with alcohol) before giving the device to another patient. It is not allowed 
to the patient to have contact with the upholstery unless different upholstery is being used for each patient. 

 

2.7

 

Safety precautions 

 

 

The device must not be overturned, thrown or dragged with the patient or without him.

 

 

 

Beside the fact that the device was made of non-flammable materials you should pay attention while closing to the 
fire sources.

 

 

 

Occurrence of corrosion points does not influence the safety but lowers the aesthetic. The manufacturer does not 
allow to use a corroded device. The upper layer guarantees safety for 24 months against corrosion.

 

 

13 

Summary of Contents for PARAPODIUM PD 180

Page 1: ...Dynamic Parapodium Model PD150 PD 180 PD 200 User s Manual 09 2016...

Page 2: ...PRELIMINARY REHABILITATION PROGRAMME FOR THE PERIOD OF ADAPTATION OF THE ORGANISM TO THE DEVICE 5 2 USER S MANUAL 6 2 1 GENERAL PRINCIPLES OF SAFETY 6 2 2 GENERAL INFORMATION 7 2 3 PREPARATION OF THE...

Page 3: ...ralysis allowing to assume upright position without additional support on crutches and sitting up Dynamic Parapodium parapodium orthosis system making active rehabilitation and movement gait possible...

Page 4: ...4 Most Common Systemic Disturbances Likely To Occur In Paraplegia WARNING The prerequisite for starting rehabilitation of the patient making use of the Dynamic Parapodium is previous consultation wit...

Page 5: ...n and inappropriate distribution of blood which leads in high temperature of the environment to the earlier occurrence of the over warming effect than in healthy subjects associated with the risk of d...

Page 6: ...round During the brakes in the previous exercise inclining and turning the trunk in various directions and or active training of the upper extremities with various loads 3 Prolongation of the period o...

Page 7: ...situations or serious injury to the user 2 2 General information The purpose of this manual The manual contains detailed information concerning the conditions of usage of the Dynamic Parapodium Model...

Page 8: ...lanes The plane perpendicular to the axis of the runners extending through the corset hanger joint after positioning the patient in the parapodium this plane extends through the axis of their hip join...

Page 9: ...unners from the ground is sufficient to initiate the movement of the parapodium The runners are equipped with lateral safety supports It is very difficult to make parapodium swing beyond the support p...

Page 10: ...ient feels safe support for the leg but without hyperextension of knee joints This ensures not only safety but also correct position of the body in the parapodium which makes walking much easier The p...

Page 11: ...relation to the axis of corset hanger joints if the patient is bent forward act according to Assembly Instructions if the patient is bent backward act according to Assembly Instructions WARNING Durin...

Page 12: ...the parapodium is assembled incorrectly act according to Assembly Instruction When the patient moves the trunk and lifts the runners from the ground the front of the runners fall down Assembly Instruc...

Page 13: ...ld appliances is acceptable Leather upholstery should be cleaned and conserved with special agents used to clean leather clothing Velour upholstery should be washed at 40 C with water containing mild...

Page 14: ...e Dynamic damages such as fissures found replace Parapodium deformations or any other visible damaged elements with damages which can affect device new ones usage or safety of the user Measure height...

Page 15: ...ease share your opinion with us and we will consider introducing the appropriate changes It is a great challenge for mdh sp z o o to create a perfect dynamic orthosis that could serve all disabled peo...

Page 16: ...24 www viteacare com Dynamic Parapodium Model PD150 PD180 PD200 User s Manual is a property of mdh Sp z o o All rights reserved Attention The devices presented in this manual are protected by appropri...

Reviews: