background image

Table 1. 

 

 

 

 

PD 150 

 

PD 180 

 

PD 200 

 

No.  Parameter (cm) 

MIN. 

 

MAX.  MIN. 

 

MAX.  MIN. 

 

MAX. 

1. 

Hip width  (S) 

 

21 

 

37 

30 

 

55 

30 

 

55 

2. 

Hip depth  (G) 

 

13 

 

25 

16 

 

30 

19 

 

33 

3. 

Height to the knees  (D) 

15 

 

61 

15 

 

74 

15 

 

89 

4. 

Height to the trochanter (K) 

55 

 

85 

70 

 

105 

85 

 

120 

5. 

Weight 

 

 

 

70 

 

90 

 

110 

6. 

Contracture of the knee (

°

 

30

° 

 

30

° 

 

30

° 

7. 

Clawfoot (

°

 

 

15

° 

 

15

° 

 

15

° 

8. 

Width of footwear 

 

11 

 

11 

 

11 

9. 

Length of footwear 

 

30 

 

 

32 

 

 

34 

10. 

Total height 

( E) 

125 

 

155 

155 

 

185 

185 

 

205 

 

 

Dynamic  Parapodium  Model  PD150,  PD180,  PD  200,  is  to  be  used  by  patients  with  parameters 
corresponding to those in Diagram 2 and table 1.

 

 

 

The device must be assembled according to “Assembly Instruction ”.

 

 

 

The  height  of  the  mounting,  the  width  and  depth  of  the  corset,  the  height  of  the  arm  brackets  and  the 
height of the knee holders as well as their span are adjusted individually for each patient and depend on 
his height, the width and depth of the hips, height of the knees, etc.

 

 

ATTENTION 

 

The distance of the corset hanger from the ground must be adjusted so that the patient, after fastening the 
back belt, is in upright position and can remain so for a desired period of time without the aid of hands. 

 

The patient must feel safe

 

 

The width and depth of the corset must be so adjusted that the patient feels kept (blocked) safely in the 
region of the pelvic girdle, but without any discomfort and unpleasant compression.

 

 

 

The  position  of  knee  holders  must  be  so  adjusted  that  the  patient  feels  safe  support  for  the  leg,  but 
without hyperextension of knee joints. This ensures not only safety but also correct position of the body in 
the parapodium, which makes walking much easier.

 

 

 

The  position  of  arm  supports  is  adjusted  to  the  patient’s  needs.  For  individuals  with  strong  arms  it  is 
recommended to fix the supports as low as possible.

 

 

 

The  device  must  be  checked  thoroughly  for  the  correctness  of  assembly  and  blockade  of  all  mobile 
elements.

 

 

 

It should be checked whether the screw blocking the elements of the stabilising rod has been fixed tight 
enough.

 

 

 

Closing and opening of the back flap and the correct functioning of the lock should be checked.

 

 

 

It  should  be  checked  whether  the  pillow  filling  the  space  between  the  buttocks  and  the  back  flap  is 
correctly fixed (if used).

 

 

In  the  parapodium  a  three-point  way  of  supporting  the  user  enables  a  firm  and  safe  grasp  of  the  patient’s  body 
eliminating  the  occurrence  of  abrasions.  The  manufacturer  allows  to  use  only  the  device  with  the  upholstery 
completely on. 

 

2.4

 

Use Of The Parapodium 

 

ATTENTION 

 

There are no contraindications for using the parapodium along with other individual orthopedic devices such as: 
orthopedic  footwear,  orthopedic  collars,  corsets  and  orthosis. 

The  decision  whether  to  apply  particular

 

orthopedic devices is always made by a physician. 

 

ATTENTION 

 

In  case  of  hyperhidrosis  and  sensitive  skin  it  is  not  recommended  to  use  the  device  without  underwear  as  a 
protection against abrasions. 

 

ATTENTION 

Dynamic

 

Parapodium

 

is  to

 

be

 

used  indoors,

 

in  the

 

range

 

of

 

temperatures

 

between

 

10

0

C

 

and

 

45

0

C

 
 
 
 
 
 

10 

Summary of Contents for PARAPODIUM PD 180

Page 1: ...Dynamic Parapodium Model PD150 PD 180 PD 200 User s Manual 09 2016...

Page 2: ...PRELIMINARY REHABILITATION PROGRAMME FOR THE PERIOD OF ADAPTATION OF THE ORGANISM TO THE DEVICE 5 2 USER S MANUAL 6 2 1 GENERAL PRINCIPLES OF SAFETY 6 2 2 GENERAL INFORMATION 7 2 3 PREPARATION OF THE...

Page 3: ...ralysis allowing to assume upright position without additional support on crutches and sitting up Dynamic Parapodium parapodium orthosis system making active rehabilitation and movement gait possible...

Page 4: ...4 Most Common Systemic Disturbances Likely To Occur In Paraplegia WARNING The prerequisite for starting rehabilitation of the patient making use of the Dynamic Parapodium is previous consultation wit...

Page 5: ...n and inappropriate distribution of blood which leads in high temperature of the environment to the earlier occurrence of the over warming effect than in healthy subjects associated with the risk of d...

Page 6: ...round During the brakes in the previous exercise inclining and turning the trunk in various directions and or active training of the upper extremities with various loads 3 Prolongation of the period o...

Page 7: ...situations or serious injury to the user 2 2 General information The purpose of this manual The manual contains detailed information concerning the conditions of usage of the Dynamic Parapodium Model...

Page 8: ...lanes The plane perpendicular to the axis of the runners extending through the corset hanger joint after positioning the patient in the parapodium this plane extends through the axis of their hip join...

Page 9: ...unners from the ground is sufficient to initiate the movement of the parapodium The runners are equipped with lateral safety supports It is very difficult to make parapodium swing beyond the support p...

Page 10: ...ient feels safe support for the leg but without hyperextension of knee joints This ensures not only safety but also correct position of the body in the parapodium which makes walking much easier The p...

Page 11: ...relation to the axis of corset hanger joints if the patient is bent forward act according to Assembly Instructions if the patient is bent backward act according to Assembly Instructions WARNING Durin...

Page 12: ...the parapodium is assembled incorrectly act according to Assembly Instruction When the patient moves the trunk and lifts the runners from the ground the front of the runners fall down Assembly Instruc...

Page 13: ...ld appliances is acceptable Leather upholstery should be cleaned and conserved with special agents used to clean leather clothing Velour upholstery should be washed at 40 C with water containing mild...

Page 14: ...e Dynamic damages such as fissures found replace Parapodium deformations or any other visible damaged elements with damages which can affect device new ones usage or safety of the user Measure height...

Page 15: ...ease share your opinion with us and we will consider introducing the appropriate changes It is a great challenge for mdh sp z o o to create a perfect dynamic orthosis that could serve all disabled peo...

Page 16: ...24 www viteacare com Dynamic Parapodium Model PD150 PD180 PD200 User s Manual is a property of mdh Sp z o o All rights reserved Attention The devices presented in this manual are protected by appropri...

Reviews: