7.2. Inspection of the Vitalograph micro
A visual inspection is recommended on a routine basis.
Examine cone and flow conditioning mesh filter for damage or
contamination. If damaged or blocked, it should be replaced with a
new part. Examine the Fleisch element and replace if damaged.
If it is suspected that the flowhead has become contaminated or
where user risk assessment identifies a need for higher level of
decontamination, then it should be cleaned as per the instructions
in ‘Cleaning and Hygiene’ on the Vitalograph website.
It is recommended that a calibration verification is carried out
following cleaning and re-assembly as recommended by the ATS/
ERS 2019 guidelines
4
.
Figure 4: Flowhead Assembly
1
Flowhead containing Fleisch
2
Flowhead Cone
4
Derived from terminology and guidance taken from ATS/ERS Standardisation of
Spirometry 2019 Update Am J Respir Crit Care Med 2019 Vol 200, Iss 8 pp e70-e88