6. The result of each stroke, expiratory (E) and inspiratory (I)
displays on the top of the screen with the number of strokes
shown in between. If they are reproducible and within 3% this
will be displayed on the top of the screen, and a syringe with
a green tick pass
will display. Pressing the forward
key will return to the main Configuration menu and the
Calibration Verification pass is recorded.
In the unlikely event the result is outside 3% or non-
reproducible, an error icon will display.
Successful verification (green syringe with a
tick).
Non-reproducible syringe strokes (red syringe
with ‘information’ in centre, after the 5th Stroke).
Outside 25% (poor technique, repeat verification
or an issue with the flowhead/device, contact
technical support).
Outside 3%, 6% (purple syringe, and shows >3%,
>6% on the top of the screen). The higher the %,
the more severe the issue.
Over 3%, repeat calibration verification.
Over 6% may indicate the device requires cleaning
or maintenance.
The following icons display after the calibration update:
Calibration update (purple syringe with purple tick
in a circle, % shown on top of screen).