Vitalograph micro - Instructions for Use
Where the user suspects that the flowhead has become
contaminated or where local risk assessment identifies a need for
higher level of decontamination, then it should be cleaned as per the
instructions in ‘Cleaning and Hygiene’ on the Vitalograph website.
Table of Cleaning/Disinfection Methods
Part
Clean/
Low Level
Disinfection
Recommended Cleaning/Low
Level Disinfection
Case
exterior
Clean
Wiping with a 70% isopropyl
alcohol impregnated cloth
provides a suitable form of
cleaning. The exposure time for
low level disinfection using 70%
IPA wipes is ≥1minute.
3
Scr
een
Clean
For the screen, lightly wipe the
surface with cotton pad or other
soft material.
NOTE: DO NOT use
any chemicals. DO NOT wipe in a
circular motion. Strokes should
be either up/down or over/back.
Where the user suspects that the flowhead has become
contaminated or where local risk assessment identifies a need for
higher level of decontamination, then it should be cleaned as per the
instructions in ‘Cleaning and Hygiene’ on the Vitalograph website.
3
Rutala, W. A. 2017 “Back to the Basics” accessed from disinfection&sterilization.org,
June 2020.