2.1 | Description of the device
(Fig.1)
Electrodes buttons
[ON/SET+]
button
[OFF/SET-]
button
Battery compartment
Electrode pad
Program name
2.2 | Key functions
[ON/SET+] Button
= Turn on the device,
device is in wait status (Diode glows
permanent); in work status you increase
intensity.
Press again the
[ON/SET+] Button
= the
device is in work status in the selected
Programm in level 1 (diode fl ashes);
Press again to increase the output intensity (to max. level 10).
[OFF/SET-] Button
in wait status = Press the button to select the program;
device is in work status = decrease intensity and turn off the device.
During operation, each press decreases the intensity by one level.
Press the button for 2 seconds to turn off the device.
2.3 | Technical information
Body material:
ABS
Channel: Single
channel
Power supply:
2x AAA alkaline batteries
Wave shape:
Biphase square-wave pulse
Pulse duration:
200 - 250 uS
Pulse frequency:
1- 80 Hz
Output voltage:
max. 35 Vpp ± 20% at 500 Ohm load
Treatment time:
30 minutes
Intensity:
Adjustable from 0 to 10
Size of electrode pad:
53 x 73 mm