8
VITA V60 i-Line
®
9 Installation and connections
9.1 Installation location
• Install the device in a dry, heated room so that the distance to the
closest wall is at least 25 cm (see section 8.2 Ambient conditions).
• When the temperature is below 15 °C (e.g. after transport), leave the
device to stand for approx. 30 minutes at room temperature before
startup.
• Ensure that the device is installed on a heat-resistant surface.
The radiation and heat development of the device is within a non-
hazardous range. However, it cannot be completely ruled out that
furniture surfaces and veneers sensitive to heat may show slight
discoloration over time due to the continuous exposure to heat.
• Avoid exposing the device to direct sunlight.
• Do not place any flammable objects near the device.
• Do not install the device in such a way that it becomes difficult to press
the main switch and pull out the power supply cable.
9.2 Device connections
USB port at the side
Connection for vacuum tube
2 device fuses*
Connection for vacuum pump
Connection for mains power supply
Main switch at the side
* The identification plates provide information about the fuses used in the device.
Fuses with other ratings may not be used.
230 V model: T 8 H 250 V
100 / 110 V model: T 15 H 250 V
Summary of Contents for V60 i-Line