55
IFU_G01079_ML_ V02202006R02
Voltage dips,
short interrup-
tions and voltage
variations on
power supply
input lines IEC
61000-4-11
0 % UT (100 % dip
in UT )
for 0.5 cycle at 0°, 45°,
90°, 135°,180°, 225°,
270°, and 315°
0 % UT
(100 % dip in UT )
for 1 cycle at 0°
70 % UT
(30 % dip in UT )
for 25/30 cycles at 0°
0 % UT
(100 % dip in UT )
for 250/300 cycle at 0°
N/A
N/A
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m, 50/60Hz
30 A/m, 50/60Hz
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commer-
cial or hospital environment.
Note: UT is the a. c. mains voltage prior to application of the test level.