This message displays in various cases - below are a list of the main cases:
CAUSES OF LO MESSAGE
DISPLAY
SOLUTIONS
Temperature reading hampered
by hair, sweat,...
Make sure there is no obstruction prior to taking the
temperature.
Temperature reading hampered
by an air flow…
Be careful not to be in the axis of an airflow, this could
cause interference with the infrared system.
Temperature reading too close
together,
ThermoFlash®
has not
had time to reset.
Please observe the minimum interval of 1 minute
between two readings - a 3 min delay is advised.
The measuring distance is too
great.
Please respect the measuring distance (between 2 and 5
cm / 0.79 - 1.97 in).