Safety Instructions
VT300 Technical Manual, Rev. A4
xi
Doc
#
TM-VT300-EN
Connection of AC Mains
Make sure that the electrical installation complies with local codes. Always connect
the AC plug to a wall socket with a protective ground.
The maximum permissible current capability of the branch distribution circuit that
supplies power to the product is 16A. The circuit breaker in the building installation
should have high breaking capacity and must operate at short-circuit current
exceeding 35A.
Always connect the power cord first to the equipment and then to the wall socket.
If the power cord cannot be readily disconnected in case of emergency, make sure
that a readily accessible circuit breaker or emergency switch is installed in the
building.
Operating Environment
Ambient
Temperature
Storage temperature: -10C to +70C (14F to 158F).
Operating temperature: -10C to +40C (14F to 104F).
Humidity
40% to 90% RH (non condensing).
Vibration
Severe vibration can affect the accuracy of weighing and damage components.
Air
The air surrounding the product should be dust-free and should not contain
corrosive gasses or other materials that could adversely effect the product.
Electromagnetic
Fields
Heavy electrical equipment should not be installed near to the weighing
apparatus.
Incoming and
Outgoing Signals
Relays and contacts connected to the equipment must have reliable and
effective interference suppression. This also applies to other equipment within
3 meters of the equipment.