Virtuox VPOD Instruction Manual Download Page 6

Vpod Instruction Manual 

 

5

 
 
 
 
 
 
 
 

CHAPTER 2 BASIC OPERATION 
2.1 Outer View 

 

Fig.1 

Description of Fig.1

 

1— Double-functional socket

for oximeter probe and data transfer cable.   

 

  Oximeter probe socket: Connect the probe with the oximeter before taking a measurement.   

— Data cable socket: Connect the oximeter with PC by a data cable for transferring data recorded 

to Virtuox software. 

2—    The 

function

 button: short press the button to change the display mode on the measurement 

screen; And on the normal screen, extended press the button for about 4 seconds to change the 

setting items. 

 

3—    The 

POWER

 on/off button   

Note

If no press on any button for 30 seconds, the oximiter will power off automatically.   

4—    The 

Setting 

button: Press this

 

button repeatedly to increase the parameter under adjustment by 

one decrement. 

5—    Display screen, displays date and time, SpO

2

 data, PR data, error information etc.   

2.2 Rear Panel   

 

Fig.2 

Install the batteries 

The oximeter can be powered by 2 AA-Size alkaline batteries (which will typically provide 50 hours of 

continuous operation), or by the optional rechargeable battery pack. 
When battery  power  is  lower  than 2.7

±

0.1V, the  sign 

 

  will  flicker  in  its display  area.  Replace  the 

battery (or rechargeable batteries) as soon as possible. The installation steps are shown as Fig. 2.3. 

 

Be  sure  to  insert  the  batteries  in  the  correct  polarity, 

as  indicated by  polarity  marking  (+  and  -) 

Summary of Contents for VPOD

Page 1: ...s of data storage High Resolution Pulse Oximetry Records at 1 second intervals Three display modes Special design for sleep monitoring LCD Displays SpO2 SpO2 waveform Pulse Rate Low power consumption...

Page 2: ...ights or patents and does not convey any license under the patent rights of our company nor the rights of others We do not assume any liability arising out of any infringements of patents or other rig...

Page 3: ...nts Basis of Operation 1 Red and Infrared ray Emitter Diode 2 Red and Infrared ray Receptor Diode 1 2 Safety Information Conception of Warning Precaution and Notice Warning Indicates a potential hazar...

Page 4: ...ry to use the SpO2 and NIBP measurement on the same arm at the same time This could potentially affect measurement accuracy Notices Operation of this device in an electromagnetic field may influence i...

Page 5: ...of operation Continuous operation Equipment not suitable for use in the presence of a flammable anesthetic mixture air or with oxygen or nitrous oxide 1 6 Intended Use The purpose and function of the...

Page 6: ...utton Note If no press on any button for 30 seconds the oximiter will power off automatically 4 The Setting button Press this button repeatedly to increase the parameter under adjustment by one decrem...

Page 7: ...as still as possible 3 2 Enter Setting mode Press the function button for about 4 second extended press the oximeter will enter into Setting mode and you will find a parameter item which include its t...

Page 8: ...h After turning on the oximeter each time the function button is pressed the oximeter will switch to another display mode shown as Fig 7 4 3 Factors that may affect the measurement During operation th...

Page 9: ...gram measurement can also provide a plethysmogram wave During this mode the PR tone is dumb Fig 7 3 SpO2 plethysmogram wave PR tone modulation Beeps in sync with the patient s pulse even under most ch...

Page 10: ...disinfecting 1 Clean or disinfect the sensor before each use 2 Unplug the sensor from the oximeter before cleaning and or disinfecting 3 Clean the external sensor surfaces and patient contact surface...

Page 11: ...ted 2 The connection between the Probe and Oximeter is loose 1 Connect the sensor 2 Please check if the probe was connected with oximeter correctly 5 5 1 Service Method a Service hours 9 00am 5 30pm M...

Page 12: ...he user s request Return Policy Warranty and non warranty returns should be handled in the following manner Contact the Technical Support Department and obtain a RMA Return Materials Authorization num...

Page 13: ...Display Type OLED double color Parameters SpO2 PR SpO2 plethysmogram waveform PR bar Mode 3 display modes Record Patient ID 1 10 Data record Up to 72 hours Data transmission Transmission method Cable...

Page 14: ...eclaration Electromagnetic Immunity for Equipment and Systems that are not Life Supporting 61000 4 6 Radiated RF IEC 61000 4 3 3V m 80Hz to 2 5 GHz d P V1 5 3 80MHz to 800MHz d P E1 5 3 800MHz to 2 5G...

Page 15: ...minimum distance between portable and mobile RF communication equipment transmitters and the Handheld Pulse Oximeter as recommended below according to the maximum output power of the communications eq...

Page 16: ...Vpod Instruction Manual 15...

Reviews: