Guidance and manufacturer's declaration - electromagnetic immunity 
 

Guidance 

and 

manufacture's 

declaration 

electromagnetic immunity 

The Handheld Pulse Oximeter is intended for use in an 

electromagnetic  environment  specified  below.  The 

customer  or  the  user  of  the  Pulse  Oximeter  should 

assure that it is used in such an environment.   

Immunity 

Test   

IEC 

60601 

Test 

level   

Compliance 

Level   

Electromagnetic 

environment 

guidance   

Electrostatic   

Discharge 

(ESD) 

IEC610004-2 

6kV 

contact 

8kV air 

6kV 

contact 

8kV air 

Floors 

should 

be 

wood, 

concrete 

or 

ceramic  tile.  If 

floor 

are 

converted with   

Synthetic 

material, 

the 

relative 

humidity  should 

be at least 30% 

Guidance and manufacturer's declaration - Electromagnetic Immunity for Equipment and 

Systems that are not Life-Supporting 

 
61000-4-6 
 
Radiated 
RF IEC 
61000-4-3 

 
 
3V/m 
80Hz 
to 
2.5 
GHz 

d= 

P

V

1

5

.

3

80MHz  to 

800MHz 

d=

P

E

1

5

.

3

800MHz 

to 2.5GHz 

Where  P  is  the 

maximum  output 

power  rating  of 

the  transmitter  in 

Watts 

(W) 

according  to  the 

transmitter 

manufacture  and 

is 

the 

recommended 

separation 

distance 

in 

meters 

(m).   

Field 

strength

from 

fixed 

RF 

transmitters, 

as 

determined  by  an 

electromagnetic 

site  survey,  should 

be  less  than  the 

compliance level in 

each 

frequency 

range. 

Inter  ference  may 

occur in the vicinity 

of 

equipment 

marked  with  the 

following symbol.       

NOTE1      At 80MHz and 800MHz, the 

higher frequency range applies. 

NOTE2    These  guideline  may  not 

apply 

in 

all 

situations. 

Electromagnetic 

propagation  is  affected  by 

absorption  and  reflection 

from structures, objects and 

people. 

a    Field  strengths  from  fixed 

transmitters, 

such 

as 

base 

 

Summary of Contents for VPOD

Page 1: ...s of data storage High Resolution Pulse Oximetry Records at 1 second intervals Three display modes Special design for sleep monitoring LCD Displays SpO2 SpO2 waveform Pulse Rate Low power consumption...

Page 2: ...ights or patents and does not convey any license under the patent rights of our company nor the rights of others We do not assume any liability arising out of any infringements of patents or other rig...

Page 3: ...nts Basis of Operation 1 Red and Infrared ray Emitter Diode 2 Red and Infrared ray Receptor Diode 1 2 Safety Information Conception of Warning Precaution and Notice Warning Indicates a potential hazar...

Page 4: ...ry to use the SpO2 and NIBP measurement on the same arm at the same time This could potentially affect measurement accuracy Notices Operation of this device in an electromagnetic field may influence i...

Page 5: ...of operation Continuous operation Equipment not suitable for use in the presence of a flammable anesthetic mixture air or with oxygen or nitrous oxide 1 6 Intended Use The purpose and function of the...

Page 6: ...utton Note If no press on any button for 30 seconds the oximiter will power off automatically 4 The Setting button Press this button repeatedly to increase the parameter under adjustment by one decrem...

Page 7: ...as still as possible 3 2 Enter Setting mode Press the function button for about 4 second extended press the oximeter will enter into Setting mode and you will find a parameter item which include its t...

Page 8: ...h After turning on the oximeter each time the function button is pressed the oximeter will switch to another display mode shown as Fig 7 4 3 Factors that may affect the measurement During operation th...

Page 9: ...gram measurement can also provide a plethysmogram wave During this mode the PR tone is dumb Fig 7 3 SpO2 plethysmogram wave PR tone modulation Beeps in sync with the patient s pulse even under most ch...

Page 10: ...disinfecting 1 Clean or disinfect the sensor before each use 2 Unplug the sensor from the oximeter before cleaning and or disinfecting 3 Clean the external sensor surfaces and patient contact surface...

Page 11: ...ted 2 The connection between the Probe and Oximeter is loose 1 Connect the sensor 2 Please check if the probe was connected with oximeter correctly 5 5 1 Service Method a Service hours 9 00am 5 30pm M...

Page 12: ...he user s request Return Policy Warranty and non warranty returns should be handled in the following manner Contact the Technical Support Department and obtain a RMA Return Materials Authorization num...

Page 13: ...Display Type OLED double color Parameters SpO2 PR SpO2 plethysmogram waveform PR bar Mode 3 display modes Record Patient ID 1 10 Data record Up to 72 hours Data transmission Transmission method Cable...

Page 14: ...eclaration Electromagnetic Immunity for Equipment and Systems that are not Life Supporting 61000 4 6 Radiated RF IEC 61000 4 3 3V m 80Hz to 2 5 GHz d P V1 5 3 80MHz to 800MHz d P E1 5 3 800MHz to 2 5G...

Page 15: ...minimum distance between portable and mobile RF communication equipment transmitters and the Handheld Pulse Oximeter as recommended below according to the maximum output power of the communications eq...

Page 16: ...Vpod Instruction Manual 15...

Reviews: