Vpod Instruction Manual
12
APPENDIX A Specifications
Notices
:
Specifications may be changed without prior notice.
The circuit diagrams, the list of components, the illustration of diagrams, and the
detailed rules of calibration, are provided exclusively to professional personnel
authorized by our company.
SpO
2
Display Range: 0-100%
Measurement Range: 70%-100%
Resolution: 1%
Accuracy: 80%-100%: ±2%
70%-79%: ±3%
0%-69%: unspecified
Heart (Pulse) Rate
Display Range: 0-254 bpm
Measurement Range: 30-235 bpm
Resolution: 1bpm
Accuracy: ±2bpm or ±2% (the larger is applied)
Alarm
Alarm: Technical alarm and physiological alarm
Modes: Visual and audible
SpO
2
: The upper limit: 100%
The lower limit: 60%
PR: The upper limit: 100bpm
The lower limit: 60bpm
Display
Type: OLED, double color
Parameters: SpO
2
, PR, SpO
2
plethysmogram waveform, PR bar
Mode: 3 display modes.
Record
Patient ID 1-10
Data record Up to 72 hours
Data transmission
Transmission method: Cable Transmission
Data Cable Interface: DB9 (Connected to Pulse Oximeter)
USB (Connected to PC)
Operation Environment
Operating temperature: 5
℃~
40
℃
Relative humidity:
≤
80%, no condensation
Atmosphere pressure: 86 kPa
~
106 kPa
Power supply: Two AA alkaline batteries
Operation time About 50 hours of typical operation
Transport and Storage Environment
Storage Temperature: -20
℃
~
55
℃
Relative humidity:
≤
93%, no condensation
Atmosphere pressure: 50 kPa
~
106 kPa
Classification according to IEC60601-1
Type of protection: Internally powered equipment
Degree of protection: Type B-Applied Part
Mode of Operation: Continuous
Safety: IEC Standard 60601-1
Dimensions:
139mmX44mmX24mm (LXWXH)
Weight:
110g (with alkaline batteries)
APPENDIX B
DB9
USB