Virtuox VPOD Instruction Manual Download Page 10

Vpod Instruction Manual 

 

9

Extended press:

 Press the button for more than 1.5 seconds. 

Long press: 

Press the button for more than 2 seconds.

 

5.3 Data replay and transmission 

The oximeter can record SpO

2

 and PR value for more than 72 hours, but the oximeter itself has no 

data replay function. You can transmit the history data to a PC using “Virtuox” software and a special 

data  cable.  As  for  detailed  setup  and  operation,  please  refer  to  the  “Virtuox”  website.     

www.virtuox.net 
Note: Virtuox can be obtained at no charge with the purchase of this device.      Please contact Virtuox 

customer service to set up an account.    877-337-7111 

5.4 Maintenance   

It is very important for user to perform daily maintenance of oximeter and parts in order to maintain its 

function  and  appearance.  Disinfection  procedures  may  be  performed  with  the  use  of  the  below 

mentioned  cleaner/disinfectants.  Failure  to  perform  these  procedures  may  result  in  invalidating  the 

warranty. Local disinfection protocols will apply. 

Please take out battery before cleaning the oximeter. 

The external surface of the oximeter can be cleaned by wiping with a damp cloth. Do not submerge 

the oximeter in any solution at any time. To do so will void the warranty 

Use the following permitted solutions: 

  Ammonia (diluted) 

  Glutaraldehyde   

  10% Bleach solution 

  Mild soapy water (diluted). 

  Do not use the following cleaners: 

 Any kind of scrubbing or scouring solution 

 Acetone   

 Alcohol-based cleaners 

Probe cleaning and disinfecting 

1.  Clean or disinfect the sensor before each use. 

2.  Unplug the sensor from the oximeter before cleaning and/or disinfecting. 

3.  Clean the external sensor surfaces and patient contact surfaces with a soft cloth moistened 

with water or a mild detergent solution. 

4.  To disinfect the sensor, wipe the surfaces with a disinfecting solution. Isopropyl alcohol is 

recommended 

5.  Do not immerse the sensor in any solution. To do so will void the warranty. 

6.  Do not sterilize the sensor by irradiation, steam autoclaving or ethylene oxide 

Battery maintenance 

  Remove the batteries if you will not be using the oximeter for a long time. 

  Charge the batteries (Ni-MH) fully prior to storage. 

  Please charge over 14 hours at first time, or will reduce the battery life. 

5.5 Troubleshooting 

a) Error Definitions 

Err 1: program memory damaged. 

Err 2: data memory damaged. 

Err 3: sensor

 

Red Emission Diode damaged. 

    Err 4: sensor Infrared-ray Emission Diode damaged.  

Err 5: sensor Infrared-ray Receipt Diode damaged. 

Err 6: exterior crystal oscillator damaged. 

Err 7: sensor emission diode or receipt diode damaged. 

Err 9: real time clock damaged. 

Err 10: EEPROM chip damaged. 

b) Possible problem and corresponding resolution 

Probl

ems 

Possible 

reason 

Solution 

SpO

2

 

or PR 

canno

t be 

displa

yed 

norma

lly 

1. Finger is not 

plugged 

correctly 

2. Patient’s 

Oxyhemoglo

bin value is 

too low to be 

measured 

1. Retry by 

plugging the 

finger 

2. Attempt 

several time 

to obtain a 

reading, If 

are sure 

that no 

problem 

exists, 

obtain 

further 

clinical 

examination 

SpO

2

 

or PR 

displa

y is 

unsta

1. Finger might 

not be 

plugged 

deep enough 

2. Finger is 

1. Retry by 

plugging the 

finger 

2. Urge the 

patient to 

Summary of Contents for VPOD

Page 1: ...s of data storage High Resolution Pulse Oximetry Records at 1 second intervals Three display modes Special design for sleep monitoring LCD Displays SpO2 SpO2 waveform Pulse Rate Low power consumption...

Page 2: ...ights or patents and does not convey any license under the patent rights of our company nor the rights of others We do not assume any liability arising out of any infringements of patents or other rig...

Page 3: ...nts Basis of Operation 1 Red and Infrared ray Emitter Diode 2 Red and Infrared ray Receptor Diode 1 2 Safety Information Conception of Warning Precaution and Notice Warning Indicates a potential hazar...

Page 4: ...ry to use the SpO2 and NIBP measurement on the same arm at the same time This could potentially affect measurement accuracy Notices Operation of this device in an electromagnetic field may influence i...

Page 5: ...of operation Continuous operation Equipment not suitable for use in the presence of a flammable anesthetic mixture air or with oxygen or nitrous oxide 1 6 Intended Use The purpose and function of the...

Page 6: ...utton Note If no press on any button for 30 seconds the oximiter will power off automatically 4 The Setting button Press this button repeatedly to increase the parameter under adjustment by one decrem...

Page 7: ...as still as possible 3 2 Enter Setting mode Press the function button for about 4 second extended press the oximeter will enter into Setting mode and you will find a parameter item which include its t...

Page 8: ...h After turning on the oximeter each time the function button is pressed the oximeter will switch to another display mode shown as Fig 7 4 3 Factors that may affect the measurement During operation th...

Page 9: ...gram measurement can also provide a plethysmogram wave During this mode the PR tone is dumb Fig 7 3 SpO2 plethysmogram wave PR tone modulation Beeps in sync with the patient s pulse even under most ch...

Page 10: ...disinfecting 1 Clean or disinfect the sensor before each use 2 Unplug the sensor from the oximeter before cleaning and or disinfecting 3 Clean the external sensor surfaces and patient contact surface...

Page 11: ...ted 2 The connection between the Probe and Oximeter is loose 1 Connect the sensor 2 Please check if the probe was connected with oximeter correctly 5 5 1 Service Method a Service hours 9 00am 5 30pm M...

Page 12: ...he user s request Return Policy Warranty and non warranty returns should be handled in the following manner Contact the Technical Support Department and obtain a RMA Return Materials Authorization num...

Page 13: ...Display Type OLED double color Parameters SpO2 PR SpO2 plethysmogram waveform PR bar Mode 3 display modes Record Patient ID 1 10 Data record Up to 72 hours Data transmission Transmission method Cable...

Page 14: ...eclaration Electromagnetic Immunity for Equipment and Systems that are not Life Supporting 61000 4 6 Radiated RF IEC 61000 4 3 3V m 80Hz to 2 5 GHz d P V1 5 3 80MHz to 800MHz d P E1 5 3 800MHz to 2 5G...

Page 15: ...minimum distance between portable and mobile RF communication equipment transmitters and the Handheld Pulse Oximeter as recommended below according to the maximum output power of the communications eq...

Page 16: ...Vpod Instruction Manual 15...

Reviews: