Operator’s Manual
Errors & Warning Messages
41/50
During the exam execution the instrument software continuously monitors all the activities performed by the various
motors actuators and pumps, by means of specific sensors.
In case some warning or error condition are detected, a message box is displayed and the user is requested to perform
some action to manage the situation Furthermore, the frontal LED will blink showing different colors depending of the
situation in which the instrument is:
o
Blue: Normal status
o
Yellow: Warning status
o
Red: Error status
4
.1
S
AMPLE OR REAGENT LIQUID LEVEL DETECTION FAILURE
The sample needle and the reagents needle are able to detect the liquid inside the samples tubes or into the reagents
bottles. If before aspirating from one of the containers the liquid level is not detected the following message box is
displayed:
The user is requested to perform one of the following actions:
Retry:
Repeat the automatic level detection.
Aspirate from the Bottom:
Force the instrument to aspirate the liquid from the bottom of the
container. Press this option only if you are sure that there is enough
liquid inside the tube or bottle.
Inspect:
It is to request the instrument to unlock the lid and allow to open it to
inspect the container and decide which action can be executed. Once
the user is able to access the container it will be possible to properly
refill it, before to issue the
Retry
command.
- Pay attention to pick the proper liquid to refill the container, do
not mix different liquids, i.e. samples and reagents!!
Invalidate:
To skip that container: this will also invalidate the programmed assays
that requires that liquid.