Operator’s Manual
Operating Procedure
16/50
2.2.
P
RELIMINARY CHECK
For the instruments and the operator safety the following conditions are required:
•
The power supply network (installation category II), must be “compatible” with the specifications of voltage and
current on the identification label affixed on the instrument side. To ensure the proper functioning of the
instrument, please check of the building electrical system output before to installation.
•
The electrical system outlet must be properly grounded as defined by the standards for outlet electrical
installation.
•
Before making the connections, verify the compatibility with the external equipment like host or PC (with the
user manual) and remember to connect the cables only with external devices power switch in off position. Verify
the continuity of the ground in the electrical system.
•
IT IS ABSOLUTELY FORBIDDEN to remove or modify or
override
any of the safety devices installed on the
instrument.
It's advised to connect the analyzer
on an uninterruptible
Power Supply system (UPS) to ensure the regular
operation in case of malfunction of the electrical system.
2.3.
P
LACEMENT
The VirClia Lotus™ analyzer is intended for use in clinical and research laboratories.
The instrument must be positioned away from any heat source, in an environment free of dust, on level surface that
is free of vibrations and oscillations on the workbench.
It’s advised to use a workbench with a sufficient range of weight tolerance and 80 cm of height to ensure an
ergonomically correct position during the loading and extracting operation from the instruments work area.
80 cm