Operator’s Manual
Operating Procedure
25/50
3.2.
B
EFORE
P
ROGRAMMING AN
A
NALYSIS
S
ESSION
Before running each session, the following operations must be carried out:
o
Turn on the switch on the right side of the instrument;
o
Turn on the Computer and make sure all the connections on the workstation are done (printer, mouse,
keyboard and video);
o
Make sure the used strips case is empty;
o
Make sure the waste tank is empty;
o
Make sure the drainpipe is placed into the specific tank or into the centralized drain if available.
3.3.
S
ESSION
P
ROGRAMMING
,
R
UNNING AND
R
ESULTS
3.3.1 W
ASH
B
UFFER
,
C
LEANING
S
OLUTION
,
AND
W
ASTE
T
ANK
Before to start the PC program make sure that the Wash Buffer Solution has been prepared following the dilution ratio
specified in the IFU.
3.3.1.1 F
ILL THE
W
ASH
B
UFFET
T
ANK
(5L
CAPACITY
)
With the prepared solution and connect the tubing: