USER'S MANUAL
Messages on display
(Rev. 7)
ENDOS AC/ACP - CE
59
•
Line voltage more than 12.1% of rated voltage
Before making an exposure the device checks the line voltage. If the
voltage is more than 12.1% compared with the rated voltage the
display shows the letters "HHH" together with the operativity green
LED which will start flashing. To reset this alert press any key
"increase" or "decrease".
•
Exposure time more than 4 seconds
In the start cycle phase (X-ray button pressed), the actual exposure
time is calculated according to the variation of line voltage. When the
time on the display is more than 4 seconds the time value will be
calculated (ES. 4.20 seconds) and simultaneously the green LED of
operational will flash. To reset this alert press any "increase" or
"decrease" key.
NOTE:
If the equipment suffers a Software crash, there is a safety timer in the
hardware that will cut off exposure after a maximum of 5.5 seconds.