USER'S MANUAL
Technical data
(Rev. 7)
ENDOS AC/ACP - CE
25
By using a tube with grid it is possible to reduce to the minimum the
time required for the anodic current, and as a consequence, the dose
adjustment to reach the steady state, so there is very little difference
between the exposure time measurement using the invasive and the non
invasive method.
Corrective actions
A practical method can be applied to get round the problem which can
be described this way:
•
In a graph you report the values of times measured using the
equipment compared with the ones displayed by the timer
(automatically corrected for the line variations): the dots of the graph
are interpolated with a straight line (if possible by the least square
method or more simply in a graphic way).
•
You determine the intercept on the Y axis of this straight line: this
can be assumed as the time value that the non-invasive device adds
to each measurement due to the radiation which reaches the device
before the anodic current is 25% of its maximum value.
•
Then you subtract this "offset" time from all the device measurements
and you proceed with comparing the time measurements displayed
by the timer.
Summary of Contents for ENDOS AC
Page 3: ...USER S MANUAL Revision history ENDOS AC ACP CE Rev 7 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 34: ...USER S MANUAL Technical data Rev 7 ENDOS AC ACP CE 29 Heat storage kJ Curve tubehead cooling...
Page 75: ...USER S MANUAL Maintenance ENDOS AC ACP CE Rev 7 70 THIS PAGE IS INTENTIONALLY LEFT BLANK...