USER'S MANUAL
Description
(Rev. 7)
ENDOS AC/ACP - CE
11
4.2.3
4.2.3
4.2.3
4.2.3
Timer
Timer
Timer
Timer
The name of ENDOS AC/ACP depends on the type of timer in use:
•
ENDOS ACP
ENDOS ACP
ENDOS ACP
ENDOS ACP
ENDOS ACP is a digital timer with microprocessor where
exposure times can be selected both manually and
automatically.
With automatic selection there is a choice of 30 pre-set times
depending on the patient’s size (small, medium or large) and
the type of tooth and acquisition mode (film, digital).
There are 33 fixed manual selection times that vary from a
minimum of 0.02 seconds to a maximum of 3.20 seconds.
The main feature of this timer is that it has an automatic
exposure time compensation for drift of nominal voltage within
±
10%.
•
ENDOS AC
ENDOS AC
ENDOS AC
ENDOS AC
ENDOS AC has the same features as the ENDOS ACP timer,
excluding automatic and digital anatomic selection.
It has manual exposure time selection only.
NOTE:
A remote X-ray button configuration can be made, outside the exam
room. This can be a pure doorbell X-ray button or a device which also
show status of the unit ("READY" and "EXPOSURE IN PROGRESS".
NOTE:
The equipment supplies two separate contacts for connection with
external signalling devices. One contact shows the status of functioning
equipment ready for use and the second one the X-ray emission.
Connection procedures and the necessary requisites for signalling
devices are given in the "Service Manual".