USER'S MANUAL
General use instructions
(Rev. 7)
ENDOS AC/ACP - CE
39
Equipment enabling key
Equipment enabling key
Equipment enabling key
Equipment enabling key
For additional safety, ENDOS AC timer is provided with the "Exposure
enabling" function.
X-rays can be emitted only after pressing the key .
The equipment enabled signal is displayed on the keyboard by switching
on the corresponding green signal. This enabling lasts for a pre-set time
(variable during setup phase between 10 and 30 seconds: default 15
seconds), after which the function itself is disabled and no X-rays are
emitted if the exposure button is touched.
NOTE:
When the READY key is pressed not only will exposure be enabled, but
the equipment will also start displaying exposure time on the display
calculated according to line voltage fluctuation compensation (see
paragraph 5.1).