Appendix A
Permissions and Settings
The Permissions and Settings on the Pbi200 are used to allow users the ability to use and set the
functions on the instrument.
Permissions
are given to the User by an Administrator or a Supervisor for various functions. These
functions are those that control certain radiation safety features (such as shutter operation) or the
handling of data (such as the ability to erase data).
Settings
on the Pb200i can be set by the User, Supervisor, or Administrator. Certain settings require
permission from the Administrator or the Supervisor to change.
Use of Permissions and Settings
•
The permissions are controlled by the Administrator and/or the Supervisor. The is accomplished
through the Password system.
•
In the most basic configuration, the Administrator (set up in the initial use of the instrument) would
configure a User and set the Permissions for that User. The Administrator can also set up the
initial settings on the instrument. This allows each user to have the initial settings be consistent
with all users. Unless restricted by permissions set by the Administrator, the User can change
the settings.
•
A less basic configuration allows the Administrator to add a Supervisor level above the User. The
Supervisor can set permissions for the User. In this configuration, the Supervisor permissions
override the Administrator permissions.
•
The User cannot change permissions. The User can change settings that are not restricted by
the permissions given. The exception to this is that the User can always change a setting to a
more restrictive option for the setting. For example, if a User is not given permission to delete
data then they cannot change that setting. If, however, they are given permission to delete data,
they can change the setting so that they cannot delete data (and change it back from there).
All permissions and settings are saved and continue with the next use of the instrument, for that User.