background image

Summary of Contents for Inami L-0955-SD

Page 1: ......

Page 2: ...cessories and product overview and structure Assembly instructions Instructions Replacing the cleaning of medical equipment consumables and storage Maintenance Troubleshooting Spec After sale service...

Page 3: ...attach ground wire Upon failure or short circuit of an electric shock Unplug the power supply plug Plug Hands must be dry with advanced plug and unplug Pointing Please An electric shock may ignite or...

Page 4: ...he position to facilitate adoption of the spring arm By 45 degrees tilting mirror to get rid of ENT Can be used such as orthopaedic surgery Microscope head and compact design for observation and made...

Page 5: ...5 objective lens 6 Filter switch 7 Microscope head rotation knob 8 Second arm balance knob 9 Second arm rotation control knob 10 Power supply unit 11 Illumination control knob 12 First arm rotation co...

Page 6: ...ling please adhere strictly to ensure that the environmental conditions at the time of use are the following environmental conditions Assembly method for the first arm based clamp L 0940SD Stabilize t...

Page 7: ...e fixed position tighten the hex screws under the one arm Screw the click in place Please make sure the swivel transformer connect the power cable Plug the power cable into the mirror cylinder of the...

Page 8: ...r using the filter switch knob on the tube side mirror In each filter set you want Type Blue retinal protection filters Color temperature change filter vertical fine adjustment knob You can refine you...

Page 9: ...oscope head is the microscope head tilt locking knob loose 0 90 up to adjust to any position you can Battery Insert the battery cable into the power supply unit and install the battery in the battery...

Page 10: ...ing The green lamp will light up when it is fully charged Lamp lighting status indicates the following meaning Yellow Lit Energized Red Lit Charging Green Blinking Practical charging completed Green l...

Page 11: ...If the dust is attached to the lens mirror please clean with a soft brush Cleaning of the lenses mirrors 3 x done remove from the zoom body Please note 10 minutes with lenses mirrors do not scratch R...

Page 12: ...ach arm To verify the movement arm is lighter and can be fixed in any position To see the movement of each arm Periodic inspection Products normally used for the 3 6 months once the monitoring On a re...

Page 13: ...orce adjustment tool tightening the locking nut adjust the microscope angle tightening force After the adjustment flat nut screws sure mirror heads rotating clamping force adjustment Fixation nut side...

Page 14: ...on Go no 2 arm balance Tightening knob is loose Case No 2 arm tightening knob Fuses and light bulb filaments to break in a short period of time Other than specified fuse or light bulb is being used Re...

Page 15: ...for the diagnosis treatment and the amount 2 Constantly monitoring equipment in General and patients without anomaly 3 Appropriate measures found abnormalities in patients and equipment patients stop...

Page 16: ...his product Since we issued the dealer purchased a warranty purchased make sure it is filled dealer taken by mark please contact Please keep this because the repair free of charge within the warranty...

Page 17: ......

Reviews: