AUTO-CUT 200 XT
A-��
APPENDIX
Manual 0-5253
Reasons why RAS Control relay would not close:
6. Pilot current flowing. There actually is pilot current flowing somewhere. Unlikely as it would normally set
the 208 fault but we have to rule it out.
a. Disconnect J41 on the Pilot board, if HF still doesn’t fire and the Relay board Pilot LED, D11, is still
on, it’s due to a fault in the detection circuits.
7. Faulty detection circuit. There isn’t any pilot current but a fault in the circuits measuring pilot current is
indicating there is current.
a. Measure between pins 8 (-) and 9 (+) on the Pilot ribbon cable from Relay board J3 to Pilot board J42. If
there is no pilot current it should be zero. Anything else indicates the Pilot board current sensor is faulty
causing the Relay board D11 to be on. Replace the Pilot board assembly.
b. If the Pilot current signal on the pilot ribbon cable was zero, measure between pins 23 (-) and 25 (+) on
the 40 pin ribbon cable between the Relay board and the CCM. This would also normally be zero if there
is no pilot current. Anything else would indicate the Relay board is faulty.
Art # 12307
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
J23
TO RELAY BOARD
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
J4
From I/O PCB
24VDC_SW
D21
GREEN
D23
RF ON
1
5
3
4
K2
RAS CONTROL
120
VA
C
120 VAC to RAS
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
J8
120
VA
C
R
ET
120 VAC_1
From J9-1
120 VAC_RET
From J9-7
CCM I/O Board
Relay & Interface Board
1
2
3
4
5
6
7
8
9
10
11
12
13
14
J59 - RAS
(98)
(99)
T2
120 / 6000 VAC
(98)
(99)
6.5K 1W
6.5K 1W
AC200XT only
(Rear Panel)
TP1
GND
TP1
GND
/ RAS ON
8. If “ /RAS ON” signal is low on pin 16 of the 40 pin ribbon cable, relative to TP1 on the CCM I/O board,
during the ignition time then we need to determine if the Relay board is defective. If /RAS ON signal is
not low the CCM or the 40 pin ribbon cable may be defective.
a. If the Relay board RF ON LED, D23, is not on while the /RAS ON signal is low, then the Relay board is
defective.
b. Is D23 is on, measure for 120 VAC on J8-3 to J8-11. If not present the Relay board is defective.
c. If 120 VAC is present at J8 during the ignition time go back and perform steps 2-4.
Summary of Contents for Auto-Cut 200 XT
Page 4: ...This page intentionally blank...
Page 8: ...TABLE OF CONTENTS...
Page 48: ...AUTO CUT 200 XT INSTALLATION Manual 0 5253 This Page Intentionally Blank...
Page 66: ...This Page Intentionally Blank...
Page 97: ...AUTO CUT 200 XT Manual 0 5253 APPENDIX A 11 This Page Intentionally Blank...
Page 107: ...AUTO CUT 200 XT Manual 0 5253 APPENDIX A 21 Display PCB Test Points TP1 GND TP2 5VDC TP3 24VDC...
Page 116: ...AUTO CUT 200 XT APPENDIX Manual 0 5253 APPENDIX 12 CAP BIAS BOTTOM PCB LAYOUT Art A 11685...
Page 117: ...AUTO CUT 200 XT Manual 0 5253 APPENDIX A 31 APPENDIX 13 CAP BIAS TOP PCB LAYOUT Art A 11686...
Page 118: ...AUTO CUT 200 XT APPENDIX Manual 0 5253 APPENDIX 14 SUPPRESSOR PCB LAYOUT Art A 11684...
Page 121: ...AUTO CUT 200 XT Manual 0 5253 APPENDIX A 35 This Page Intentionally Blank...
Page 163: ...AUTO CUT 200 XT Manual 0 5253 APPENDIX A 77 This Page Intentionally Blank...
Page 165: ......