background image

2-8.REQUENCY

F

 

RANGE 

ACCURACY 

RESOLUTION 

10Hz 

±(0.5%+4) 

0.01Hz 

100Hz 

0.1Hz 

1000Hz 

1Hz 

10kHz 

10Hz 

100kHz 

100Hz 

1MHz 

1kHz 

60MHz 

10kHz 

Input sensitivity:0.7V   

Over load protection: 250V DC or AC peak value; 

2-9.Transistor(hFE) 

RANGE 

  VALUE 

TESTING CONDITION 

NPN OR PNP 

0

1000 

Basic current approx.15μA

Vce approx.4.5V 

2-10. Diode and continuity performance test 

range 

VALUE 

TESTING CONDITION 

   

   

Forward voltage drop of diode 

Forward  DCA  is  approx.  0.5mA, 
the  backward  voltage  is  approx 
1.5V 

Buzzer  makes  a  long  sound  while 
resistance is less than (50±10)Ω 

Open  circuit  voltage  is  approx. 
0.5V 

Over load protection: 250V DC or AC peak value; 

CAUTION

DO NOT INPUT VOLTAGE AT THIS RANGE! 

2-10. TEMPERATURE

(℃)

 

RANGE 

VALUE 

RESOLUTION 

-40

-1000

 

400

 ±(0.8%+4) 

≥400

 ±(1.5%+15) 

1

 

0F-1832

 

750

 ±(0.8%+5) 

≥750

 ±(1.5%+15) 

 

Sensor

TP01(K type thermocouple) 

CAUTION

DO NOT INPUT VOLTAGE AT THIS RANGE! 

4. OPERATION 

4-1. Panel description 

1. LCD: display the measuring value and unit. 

2.Function key: 

2-1.HOLD  key

press  it,  the  presently  measured  value  is  held  on  LCD  and  HOLD  symbol 

displays. Press it again, HOLD symbol disappears, and the meter is exited the holding mode. 

2-2. REL key

press it

reading clear

turn into relative value measurement states

“REL”symbol 

displays, Press it again,“REL” symbol disappears, and the meter is exited the relative mode. 

2-3.  Hz/DUTY  key

When  measuring  the  AC  Voltage  (Current)

press  it

it  will  switch 

Frequency/duty  cycle/Voltage(Current)

When  measuring  the  Frequency  ,it  will  switch 

frequency/duty cycle

1~99%

 

2-4.”DC/AC”key: switch DC and AC work mode.. 

2-5.RANGE key

select auto range or manual range mode, Auto range is the original states, it 

will display ”AUTO” symbol 

press it change to manual range .Press it more than 2 second , it 

will return to auto range states 

2-6.MAX/MIN  key  :press  it,  turn  into  MAX  mode,  it  will  hold  the  max  value  of  measuring, 

press it again ,turn into MIN mode, it will hold the min value of measuring. No auto power off 

and  analog  bar  display  under  this  mode.  Press  it  more  than  2  second  ,  it  will  exit  MAX/MIN 

states. 

3. hFE transistor COM                       

4. Knob

Switch measuring function and range. 

5. Temperature COM 

6. Voltage

Current

resistance

frequency and GND 

COM. 

7.battery case. 

See picture 1 

4-2.DCV    measuring 

1.Select the knob to 

  range. 

2. Insert the black test lead to “COM” terminal and the red one to “V/Ω/Hz” terminal. 

3.Auto  range  is  the  original  states,  it  will  display  ”AUTO”  symbol

press  “RANGE”  key 

change to manual range mode

600mV

6V

60V

600V

1000V range is selective; 

4. Connect the leads crossly to the electric circuit under test; LCD displays polarity and voltage 

under test connected by the red test lead. 

Note

 

1. Firstly users should select the knob to the highest range, if users had no idea about the range 

of  voltage  under  test,  and  then  select  the  proper  range  based  on  displaying  value.  If  LCD 

displays “OL”, it means meter is over the max. Value of this range, thus should select the knob 

to a higher range. 

2. Do not input a voltage over DC 1000V. 

3. Be carefully while measuring a high voltage. DO NOT touch the high voltage circuit. 

4.When the measuring voltage large than DC1000V 

the built –in buzzer will be sounds. 

4-3.ACV measuring 

1.Select the knob to “

” range

 

2. Insert the black test lead to “COM” terminal and the red one to “V/Ω/Hz” terminal. 

3.  Auto  range  is  the  original  states,  it  will  display  ”AUTO”  symbol

press  “RANGE”key 

change to manual range mode

6V

60V

600V

750V range is selective; 

4. Connect the leads crossly to the electric circuit under test, LCD displays voltage by the test 

lead. 

Note: 

1. Firstly users should select the knob to the highest range, if users had no idea about the range 

of  voltage  under  test,  and  then  select  the  proper  range  based  on  displaying  value.  If  LCD 

displays “OL”, it means meter is over the max. Value of this range, thus should select the knob 

to a higher range. 

2

.

 Do not input a voltage over AC 750V. 

3. Be carefully while measuring a high voltage. DO NOT touch the high voltage circuit. 

4.When the measuring voltage large than AC750V 

the built –in buzzer will be sounds. 

4-4.DCA measuring

 

1.Insert  the  black  test  lead  to  “COM”  terminal  and  the  red  one  to  “mA”  terminal  (the  Max. 

600mA) or to “20A”(the Max.20A); 

2.Select the knob to a proper DCA range, press “DC/AC” key to select the measurement mode, 

then  connect  the  leads  crossly  to  the  electric  circuit  under  test;  LCD  displays  polarity  and 

current under test connected by the red test lead. 

Note: 

1.

 

Firstly users should select the knob to the highest range, if users had no idea about the 

range of current under test, and then select the proper range based on displaying value . 

2.

 

If the LCD displays “OL”

it means the current is over range. Now you need to select the 

knob to the higher. 

3.

 

Max. input current is 600mA or 20A

subject to where the red test lead insert to

, too 

large current will damage the fuse.   

4-5.ACA measuring 

Summary of Contents for VC99

Page 1: ...relative humidity 80 1 8 Store environment 10 40 relative humidity 80 1 9 Power supply 2pcs 1 5V battery AAA 7 battery 1 10 Dimension 185 93 35mm length width height 1 11 Weight approx 290g including battery 1 12 Accessories test leads user manual temperature probe holster gift box and 2 1 5V battery 2 TECHNICAL FEATURES 2 1 Accuracy a reading data digits environment temperature at 23 5 relative h...

Page 2: ... terminal 3 Auto range is the original states it will display AUTO symbol press RANGE key change to manual range mode 600mV 6V 60V 600V 1000V range is selective 4 Connect the leads crossly to the electric circuit under test LCD displays polarity and voltage under test connected by the red test lead Note 1 Firstly users should select the knob to the highest range if users had no idea about the rang...

Page 3: ...econds 4 8 Frequency measurement 1 Insert the black lest lead to COM terminal and the red one to V Ω Hz terminal 2 select the knob to Hz range insert the test leads or shielded cable to the signal source or the load which is tested 3 Press Hz DUTY key to switch frequency duty cycle it will display the frequency or duty cycle of the signal source which is tested Note 1 There is only the auto range ...

Page 4: ...ol 8 If do not operate for a long time should take out the battery to avoid leakage 9 When symbol displays should replace the battery following the steps 9 1 see picture 2 move holster at first 9 2 Unlock the button and remove the battery case 9 3 Take out the old battery and replace the new one It s better to use alkaline battery for longer life 9 4 fit on battery case and bolt 9 5 see the pictur...

Reviews: