Adjusting the Scale
-48-
4-2 Testing the Balance
The amount of offset with respect to the reference weight can be checked. This is called “span
test.”
Performing a span test will not calibrate the balance.
参 考
Use a weight for calibration that weighs 50% of the weighing capacity or heavier. To calibrate
more accurately, use a weight that is equivalent to the weighing capacity.
Please contact us if you wish inquire about or place an order for calibration weights.
1
Set to the function setting mode.
(Refer to “Section 2-5: Function Setting Basics (P.
16).)
Press and hold the [Function] key. After “
”
is displayed, release the finger.
2
Set to the span test mode.
Press the [Function] key several times to select “
” .
Press the [Zero/Tare] key several times to select
“
” .
3
Save the setting.
Press the [Set] key.
The balance goes back to normal display.
4
Start span test.
<Press and hold>
Keep holding down the [Function] key until “
” is displayed.
After zero-point testing (“
” blinks), the display
changes to “
” .
5
Place the weight on the pan.
After “
” is displayed, place the weight on
the pan.
The test starts.
6
The error is displayed.
The display first indicates “
” and then the
weight error of the balance.
A weight error is as follows:
Weight error = True value – Current weight
When a weight error is more than “0”, the displayed
weight is less than the actual weight.
Press any key to return to normal display.
Reference