3-6 Adding Multiple Measurements
-35-
Weighing with addition function
1
Place the first sample.
After an asterisk (*) is displayed, press the [Set]
key.
The measured value is saved and a sigma [
Σ
] sign is
displayed for several seconds.
2
Place another sample (cumulative).
Place additional samples (net addition).
After an asterisk (*) is displayed, press the [Set]
key.
The measured value is saved, and a [
Σ
] sign and
the cumulative weight are displayed for several
seconds.
Repeat this operation to weigh all the samples to
be summed.
注 意
After unloading the previous sample, check that the
display indicates “0,” and place the next sample.
参 考
Cumulating weight is also possible as follows: Press
the [Zero/Tare] key without unloading the balance and
then place next samples.
3
Display the cumulative weight.
<Press twice>
Press the [Function] key twice.
A [
Σ
] sign and the cumulative weight are displayed.
参 考
-
Pressing the [Zero/Tare] key when a cumulative weight is displayed clears the cumulative
weight.
-
You can add samples when an asterisk (*) is displayed.
- When
“
” is displayed by pressing the [Set] key, it indicates that you put
additional samples on twice or that you unloaded some samples.
Caution
Reference
Reference