14
4.2 Description of Functions (cont.)
[ ] (tl_Hong Kong)
[
Upper right] (tl_Singapore,Malaysia)
[
Lower right] (tl_Taiwan)
[
mom
]
[ ] (to)
☆
3
[
Pcs
]
☆
4
[
%
]
Setup of
measurement units
to be displayed
※
1
Register
measurement units
by selecting
Function key
☆
5
Unit not set
☆
Disable
GLP-compliant
print
※
2
Enable
Prints a GLP-compliant format at calibration.
(Refer to “9.3 GLP-compliant print” (Page 31).)
☆
1
〜☆
5: default factory settings [
]
〜
[
]
※
1 [
] cannot be set at [
].
※
2
This setting is only available with the AJH-E Series.
4.3 Interface Section
Displayed when [
] is set to [
] or [
]
Item Set
Value
Description
Stop output
Output continuous at all times
Output continuous if stable (stop output if unstable)
Outputs once by pressing Memory key (irrespective of
whether stable).
Outputs once if stable. Outputs if the balance is stable
when a sample is loaded after the preceding sample has
been removed and the balance indicated zero, or less.
Outputs once if stable, and stops output when unstable.
Even if the sample is not replaced, the balance is output
once when it stabilizes next time (including the zero
indication).
Outputs once if stable, and outputs continuously when
unstable. Even if the sample is not replaced, output of
the balance stops when it stabilizes after being output
once.
Output Control
☆
Pressing Memory key causes the balance to output
once when stable.
☆
1200 bps
2400 bps
4800 bps
Baud Rate
9600 bps
☆
None
Odd
Parity
Even
Displayed only when [
(7-digit numeric format)] is specified.
☆
denotes a factory-setting
※
The data interval in continuous output mode is 0.1 to 1 second. (The interval varies depending
on weighting conditions and other factors.)