
OLP-85/OLP-85P
7
O
PERATING
MANUAL
UPDATE
1 I
NTRODUCTION
Connector type
The connection is made with an universal push pull (UPP)
adapter for all 2.5 mm connectors (UPP adapter for 1.25 mm
ferrules are also available). The OLP-85 works with a .../PC and .../
APC connector as well.
Powermeter mode
Both modulated and unmodulated light signals can be
measured. The average power of modulated light signals is
displayed. The power meter is calibrated at 850 nm, 980 nm,
1310 nm, 1490 nm, 1550 nm and 1625 nm.
The “Auto-
” function allows automatic wavelength detection.
This application requires a wavelength encoding light source,
e.g. one of the OLS-85 Optical Light Sources.
Operating manual update
Continuing enhancement and further development of the
SmartClass™ Fiber family may mean that this operating manual
does not cover all the latest functions of your instrument.
If the operating instructions about features provided by your
instrument are missing, please visit the Viavi web site to check if
additional information is available.
To download the latest user manual:
1.
2.
Select your SmartClass™ Fiber model from the product line.
3.
Open the download area and download the latest user
manual.
Summary of Contents for SmartClass OLP-85 Series
Page 82: ...OLP 85 OLP 85P 82 ROHS RoHS ...
Page 83: ...OLP 85 OLP 85P 83 ROHS ...