For 3 jumpers dual-ended,
1.
Connect FC1 to the Local TX port and SC1 to coupler 1.
2.
Connect SC5 to the Local RX port and SC6 to coupler 2.
3.
Connect FC2 to the Remote TX port and SC2 to coupler 4.
4.
Connect SC3 to the Remote RX port and SC4 to coupler 3.
5.
Connect your jumper’s SC end SC7 to coupler 1 and SC8 to coupler 3.
6.
Connect your jumper’s SC end SC9 to coupler 2 and SC10 to coupler 4.
Please ensure the fiber cables are cleaned using the cleaning kit provided in the kit.
TX
FC1
SC5
RX
TX
FC2
SC3
RX
SC1
SC6
SC7
SC9
SC8
SC10
SC4
SC2
7.
Press the SETUP button > Testing Settings > Reference Method and select “Three
Jumper” in the Remote Unit.
8.
Press the TOOLS button > Set Reference and press the “ok” button to begin set
reference.
9.
Check that the Set Reference result is ≤0.15 dB before proceeding. Clean or replace
reference cords if necessary, and set reference again if value exceeds 0.15 dB.