INSTALLATION REQUIREMENTS & CAUTIONS 

 

1.

 

The Model 359 is shipped from the factory with a protective cover over the sensing diaphragm.  Care must be taken at all times to prevent 

the diaphragm from becoming damaged or ruptured. 

2.

 

The Model 359 is shipped with a silicone (FDA Approved) O-Ring that provides a seal to the tank receiver or adapter.  Care should be taken to 

ensure the O-Ring is seated properly.  If it is not, product will leak out of the breather passageway.  A part number for a spare O-Ring is listed 

in the accessories section of this Installation Manual. 

3.

 

The threaded tank shell-retaining nut, optionally available from Viatran, secures the transmitter into the tank receiver or adapter.  It is 

recommended that this nut be hand tightened to provide the required force to seat the O-Ring.  An optional wrench is available for this nut if 

solidified product prevents the nut from being loosened by hand.  When the transmitter is removed from the receiver or adapter, the O-Ring 

should be removed, inspected and cleaned or replaced if necessary. 

4.

 

It is important that breather passageways be kept clear of solidified product.  If these holes become blocked, the transmitter output will be 

erratic due to changes in barometric pressure.  A probe, such as a pipe cleaner, may be used to clear the reference vent holes.  If the 

problem persists, soak the unit in Isopropyl Alcohol to dissolve any blockage. 

5.

 

If the transmitter must be removed for an extended period of time, a dummy plug is optionally available to seal the tank so tank or system 

can be put back in service.  Like the Model 359, the plug is secured into the tank receiver or adapter by the retaining nut and sealed by the O-

Ring.  For the appropriate length plug, please contact one of our Application Engineers. 

6.

 

This product cannot be used in sterilization systems as described in the 3A Sanitary Standard 74 – XX (latest rev), Sections E1.9. 

7.

 

During product installation, the customer shall verify that all o-rings, gaskets, or seals meet or exceed the requirements of the 3A Sanitary 

Standard 74 - XX (latest rev). 

8.

 

This product utilizes a sensing diaphragm that is “corrugated” or “grooved”, and shall be mounted at a minimum of 10 degrees to meet the 

intent of 3A Sanitary Standard 74 – XX (latest rev), for self draining when installed in a process tank or piping by the customer.  

9.

 

The mounting for the Model 359 Tank Spud, shall have the “flats” of the shell aligned so as to allow for leak detection of the tank mounting. 

10.

 

This product utilizes a fill fluid, DC-200 (CAS# 63148-62-9), and has been determined to be non-toxic.  

11.

 

 Adapters for Model 359, in general, are not 3A approved.  Customer must ensure that the installation meets the requirements of 3A             

Sanitary Standard 74 – XX (latest rev) for both leak detection and self draining when installed in a process tank or piping by the customer. 

 

 

 

BASIC OPERATION & INTERNAL CALIBRATION CIRCUIT 

 

Note: 

 The casting of the module has raised numbers at the adjustment screw positions.  Those numbers reflect the following function of the 

assigned screw. 

 

 

Screw 1:  Top Left, Coarse Zero Pot (CCW - Lower, CW - Raise) 

 

Screw 2:  Top Right, Fine Zero Pot (CCW - Lower, CW - Raise)  

 

Screw 3:  Lower Left, Internal Cal. Activation Pot (12 or 6 o’clock position – Disengage;   3 o’clock position  - Engage)    

 

Screw 4:  Lower Right, Span Adjustment Pot (CCW - Lower, CW - Raise) 

 

1.

 

Outside of using the calibration circuit or the availability of a metrology lab for obtaining a pressure (Full Scale) reading, no object should be 

used to physically move the sensor.  The diaphragm can easily be ruptured by a fingernail.  

2.

 

Once proper connections and power are applied, the unit should begin to function. 

3.

 

Check the zero (4.00 mA) with the performance certificate to verify condition.  If it is out of specification from the performance certificate, 

use the coarse and fine zero adjustment screws to obtain the proper reading. 

 

4.

 

With 

no

 pressure applied, engage the calibration circuit by slowly rotating the calibration screw until the current output rises to about 8.80 

mA’s (for a standard calibrated range). Check the reading with the performance certificate.  If the reading is off, adjust the span screw to 

raise or lower the reading – return the calibration circuit to the original position. 

5.

 

When making zero and/or span adjustments, there may be some interaction between the zero and span readings - repeat adjustments if 

needed.  

6.

 

Always return the calibration screw to off or full CCW position before installing the transmitter into the application.  

Note:  This is a common 

reason for incorrect readings at start-up.  The calibration circuit must be deactivated in order for the product to perform properly.

 

7.

 

Once the product has been tested and functionality verified, the power and meter connections may be removed and the product reconnected 

to the instrumentation loop. 

 

 

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for 359

Page 1: ... F 3 21 17 MODEL 359 SANITARY PRESSURE TRANSMITTER INSTALLATION MANUAL 199 Fire Tower Drive Tonawanda NY 14150 Toll Free 1 800 688 0030 International 716 629 3800 Fax 716 693 9162 www viatran com solutions viatran com ...

Page 2: ...e certificates that are shipped with the product These documents provide important information on the product s calibration operation safety precautions recommended maintenance re calibration requirements repair service instructions and warranty information These documents are updated from time to time as changes to the product occur and should be reviewed at receipt so that proper and safe instal...

Page 3: ...d in a process tank or piping by the customer 9 The mounting for the Model 359 Tank Spud shall have the flats of the shell aligned so as to allow for leak detection of the tank mounting 10 This product utilizes a fill fluid DC 200 CAS 63148 62 9 and has been determined to be non toxic 11 Adapters for Model 359 in general are not 3A approved Customer must ensure that the installation meets the requ...

Page 4: ... Tank Adapters Tank Shell Receiver Tank Shell Bezel Tank Plugs PREVENTATIVE MAINTENANCE Note This checklist should be followed each time a transmitter requires preventative maintenance Depending on product usage cycle all maintenance should be done on a schedule that is reflective of the application 1 Cleaning in place is the recommended process that should be performed as the system or process pe...

Page 5: ...s perceived to be exhibiting problems it can be returned to Viatran for analysis and or repair It is suggested that only Viatran personnel attempt repair of the product Any damage resulting from customer disassembly would result in a loss of coverage under the warranty policy All Viatran products are able to be repaired at minimal cost if simple cleaning and precautions are taken in the handling a...

Reviews: