Select
button to start a measurement again.
Select
button to review waveform.
Press
button to open the help information.
In the Result Screen, if there is no operation for 1 minutes, the device will automatically return
to upper menu until it powers off.
1.6 Symbols
1.7 Abnormal ECG
Display content
Conditions
High HR(Fast heart rate)
More than 100/ min
Low HR(Slow heart rate)
Less than 60/ min
Irregular beat
No R-waves detected for 2 seconds and more
R-R variation rate is more than 10% of the normal
intervals
R-R interval is less than 80% of the normal intervals
When PVC detected
Regular beat
None of the above
2.
Getting Started
2.1 Unpacking
Before unpacking, examine the packing case carefully for signs of damage. If any damage is
detected, contact the carrier or us. If the packing case is intact, open the package and remove the
equipment and accessories carefully. Check all materials against the packing list and check for any
mechanical damage. Contact us in case of any problem.
Warnings
Save the packing case and packaging material as they can be used if the device must be
reshipped.
Keep the warranty card, which is useful within the period of warranty.
When disposing of the packaging material, be sure to observe the applicable waste control
Symbol
Meaning
Application part type BF
Manufacturer
Symbol for “ENVIRONMENT PROTECTION – Waste electrical products
should not be disposed of with household waste. Please recycle where
facilities exist. Check with your local authority or retailer for recycling
advice”.
IP22
Against ingress of solid foreign objects ≥12.5mm diameter, against
dripping (15° tilted)
MRI unsafe, presents hazards in all MR environments as device contains
strongly ferromagnetic materials.
Follow operating instructions
Non-ionizing radiation
Battery indicator