16
Surge
IEC61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Voltage dips, short
Interruptions
and
Voltage variations
on power supply
input lines
IEC61000-4-11
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user
of our product requires
continued operation
during power mains
interruptions, it is
recommended that our
product be poweredfrom
an uninterruptiblepower
supply or abattery.
Power
frequency
(50/60 HZ) magnetic
field IEC 61000-4-8
3 A/m
3 A/m
Power
frequency
magnetic fields should be
at levels characteristic of a
typical location in a typical
commercial or hospital
environment.
Note: U
T
is the AC mains voltage prior to application of the test level.
Guidance and Declaration - Electromagnetic Immunity
The Health Monitor is intended for use in the specified electromagnetic environment. The
customer or the user of the Health Monitor should assure that it is used in such an environment
as described below.
Immunity
test
IEC60601
testlevel
Compliance
level
Electromagnetic
environment
-
guidance
Conduced RF
IEC61000-4-6
3 Vrms 150
kHz to
80 MHz
outside ISM
bands
3 Vrms 150 kHz
to
80 MHz
outside
ISM
bands
Portable and mobile RFcommunications
equipment should be used no closerto
any part of the system, including cables,
than the recommended separation
distance calculated from the equation
appropriate for the frequency of the
transmitter. Recommended separation
distances:
d
1 .2
P