Caution:
In case of verified scales in II OIML class with divisions e=10d
(where: d – reading division, e – verified division) the last digit
will be marked as shown below:
7. BALANCE LEVELLING
x
Operation temperature range for this device is outlined as
+15
q
C
y
+30
q
C;
x
After powering up this device requires 30 minute worming up;
x
During the worm-up time the indication can change;
x
User calibration should be performed after the warm-up time.
x
Temperature and humidity changes during operation can increase
measurement errors, which can be minimized by performing the
user calibration process.
8. KEYPAD
Keypad of WLC/A2 series
Keypad of WLC…/C/2 series
- 11 -