x
Message
<nImh>
appears on the display (for scales
of WLC C1…C2, WLC/B1 series) or
<SLA>
(for scales
of WLC/A2, WLC…/C/2 series),
x
In case of damaging accumulators or lack of it the
pictogram
blinks quickly (period about 0.5 sec).
Notice:
WLC C1…C2 and WLC/B1 scales are equipped with
NiMH
batteries
packs,
R6 (AA)
size and Power adapters for supplying from mines.
Procedure:
x
Enter the submenu
<P5.othr>
and then:
CHr6 YES
- enabled
CHr6 no
- disabled
Return to weighing:
See 11.2.2.
15.4.4. Formatting rechargeable battery packs
WLC C1…C2 and WLC/B1 scales are equipped with
NiMH
batteries
packs,
R6 (AA)
size and Power adapters for supplying from mines.
They need formatting after first powering up. It is crucial for batteries
lifetime to undertake this process. Formatting consist in charging and
total discharging (without meantime charging).