at same time 3seconds
〖
P0000
〗
Second digit flash
,
use function key to
move digit, set value.
Press
【
TARE
】【
ZERO
】【
CLEAR
】
〖
P1358
〗
Input 1358
,
1358 is calibration password
Press
【
SHIFT
】
〖
SCALE
〗
Confirm
,
into calibration program
Press
【
SHIFT
】
〖
CALSP
〗
No load
,
the screen display 0,and wait
until STB and ZERO light on.
Press
【
SHIFT
】
〖
LoAd1
〗
Waiting load
___________________________________________________________________________
____
Hanging standard weight
5000kg
Waiting until STB light on.
Press
【
SHIFT
】
〖
05000
〗
Display capacity you choose when
parameter setting
,
and first digit flash
,
Do wait until STB light on.
Press
【
SHIFT
】
〖
-------
〗
Confirm
〖
-oL-
〗
〖
End
〗
Calibration end
〖
5000.0
〗
Display calibrated capacity , sometimes it
display 4999.6,4999.8,5000.2,5000.4
Press
【
TARE
】
〖
05000
〗
Move decimal position
Put down weight
〖
0
〗
Check zero point
,
it displays 0.0, and
STB, ZERO light on.
Hanging weight 5000kg
again
〖
5000
〗
Below is test program to check scale
accuracy.
Put down 500kg weight
and hang on 40kg weight
〖
40
〗
40kh is minumum apacity
(
20e
)
, 50kg,
100kg is also acceptable
Put down 40kg weight.
Hang on 1000kg weight.
〖
1000
〗
1000kg is 500e, check accuracy
Put dowm 100kg weight
〖
0
〗
Check zero point
Generally, if you have full capacity standard weight, you are strongly recommended to
use this progress, only when you find the linearity is not good, so you have to choose
following progress.
2
.
Three points calibration program
OPERATION
DISPLAY
ILLUSTRATION
〖
0.0
〗
Take Cap:1000kg scale as example
,
Pay
attentionthat the screen display 0 after
hanging tare weight, you can hang tare
weight at first, then turn on.